Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 99291: Ferno-Washington Inc: Power F2 Powered Fastening System

FDA recall 99291 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Product may experience accelerated roller wear and potential deformation due to interference between the roller assembly and rail. In the worst-case condition, rail mechanism may become lodged during operation and movement of the stretcher along the rail may no longer be possible through either powered or manual operation. May cause delays in transport of patient to healthcare facility.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

271 more codes are in the record below.

  • Power F2 Powered Fastening System, Product No. 0567091 Powered fastening system intended for professional use only. Designed to load, unload, and secure a compatible wheeled stretcher inside a ground-based ambulance. Item no. 0567091

    Lot or code
    UDI-DI 00190790006570 Serial Numbers: 25C227687 25C228204 26C232382 26C233091 25C227688 25C228205 26C232383 26C233092 25C227689 26C231797 26C232384 26C233093 25C227690 26C231798 26C232709 26C233094 25C227691 26C231799 26C232710 26C233095 25C227692 26C231800 26C232711 26C233096 25C227693 26C231801 26C232712 26C233097 25C227694 26C232002 26C232713 26C233098 25C227695 26C232004 26C232714 26C233099 25C227696 26C232005 26C232715 26C233100 25C227697 26C232006 26C232716 26C233111 25C227698 26C232007 26C232717 26C233112 25C227699 26C232008 26C232718 26C233113 25C227700 26C232009 26C233032 26C233114 25C227701 26C232010 26C233033 26C233115 25C227702 26C232011 26C233035 26C233116 25C227703 26C232374 26C233036 26C233117 25C227704 26C232375 26C233037 26C233118 25C227707 26C232377 26C233038 26C233119 25C227708 26C232378 26C233039 26C233120 25C227709 26C232379 26C233040 26C233443 25C227710 26C232380 26C233041 26C233444 25C227711 26C232381 26C233090 26C233446 26C233448 26C234080 26C234360 26C234647 26C233449 26C234081 26C234361 26C234648 26C233450 26C234082 26C234362 26C234649 26C233451 26C234083 26C234363 26C234797 26C233452 26C234084 26C234364 26C234798 26C233692 26C234085 26C234566 26C234799 26C233693 26C234086 26C234567 26C234800 26C233694 26C234087 26C234630 26C234801 26C233695 26C234088 26C234631 26C234802 26C233696 26C234089 26C234632 26C234804 26C233697 26C234090 26C234633 26C234805 26C233698 26C234345 26C234634 26C234806 26C233699 26C234346 26C234635 26C234807 26C233700 26C234349…
    FDA recall number
    Z-2760-2026
    Quantity
    278

    The text above is the beginning of the record; 1,310 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, GA, IA, ID, IL, IN, KY, LA, MD, ME, MN, MO, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Poland, Netherlands, Canada, Germany, Norway, United Kingdom, Denmark and New Zealand.

Source

Issued by
U.S. Food and Drug Administration
Firm
Ferno-Washington Inc
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Ferno-Washington Inc, Month: July 2026, Week 2026-W31. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Product may experience accelerated roller wear and potential deformation due to interference between the roller assembly and rail. In the worst-case condition, rail mechanism may become lodged during operation and movement of the stretcher along the rail may no longer be possible through either powered or manual operation. May cause delays in transport of patient to healthcare facility.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 99291 on July 29, 2026.