What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
392 more codes are in the record below.
OmniTract Surgical Retractor System, which includes 633 of the firm s product codes.
- Lot or code
- Product Code 3032, UDI 10381780268468; Product Code 3055, UDI 10381780268505; Product Code 3069, UDI 10381780268512; Product Code 3076, UDI 10381780268529; Product Code 3086, UDI 10381780268536; Product Code 3092, UDI 10381780268550; Product Code 3093, UDI 10381780268567; Product Code 3096, UDI 10381780268581; Product Code 3098, UDI 10381780268598; Product Code 3099, UDI 10381780268604; Product Code 3133, UDI 10381780268680; Product Code 3137, UDI 10381780268697; Product Code 3142, UDI 10381780268710; Product Code 3146, UDI 10381780268734; Product Code 3148, UDI 10381780268758; Product Code 3176, UDI 10381780268826; Product Code 3177, UDI 10381780268833; Product Code 3185, UDI 10381780268857; Product Code 3201, UDI 10381780268871; Product Code 3203, UDI 10381780268888; Product Code 3216, UDI 10381780268895; Product Code 3217, UDI 10381780268901; Product Code 3218, UDI 10381780268918; Product Code 3219, UDI 10381780268925; Product Code 3220, UDI 10381780268932; Product Code 3274, UDI 10381780268994; Product Code 3312, UDI 10381780269038; Product Code 3335, UDI 10381780269045; Product Code 3336, UDI 10381780269052; Product Code 3342, UDI 10381780269076; Product Code 3343, UDI 10381780269083; Product Code 3391, UDI 10381780269168; Product Code 3398, UDI 10381780269212; Product Code 3424, UDI 10381780269243; Product Code 3426, UDI 10381780269250; Product Code 3483, UDI 10381780269267; Product Code 3485, UDI 10381780269274; Product Code 3520, UDI 10381780282273; Product Code 3529,…
- FDA recall number
- Z-2714-2026
- Quantity
- 633 units
The text above is the beginning of the record; 23,976 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Distribution (FDA): Worldwide distribution - US Nationwide and the countries of Australia, Austria, Azerbaijan, Bangladesh, Belgium, Brunei Darussalam, Bulgaria, Canada, Chile, China, Czechia, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Korea (the Republic of), Kuwait, Lithuania, Malaysia, Malta, Mauritius, Mexico, Nepal, Netherlands (the), New Zealand, Norway, Pakistan, Palestine, State of, Philippines (the), Poland, Portugal, Puerto Rico, Qatar, Republic of North Macedonia, R¿union, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan (Province of China), Thailand, Trinidad and Tobago, Turkey, United Arab Emirates (the), United Kingdom of Great Britain and Northern Ireland, (the), Viet Nam.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Integra LifeSciences Corp. (NeuroSciences)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.