What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
121 more codes are in the record below.
GEM Premier 5000 with IQM2 Model/Catalog Number: 00024019255 and GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Portable blood analyzer system. Affected Accessory: Bar Code Reader (BCR) Wand Part No. 00024015859
- Lot or code
- GEM Premier 5000 Part number: 00024019255, UDI: 08426950870227 GEM Premier 7000 Part number: 00000015279, UDI: 0195226000924 Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634
- FDA recall number
- Z-2821-2026
- Quantity
- 122
Brand Name: GEM Premier 7000 with iQM3 Product Name: GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Intelligent Quality Management 3 (iQM3) is used as the quality control (QC) and assessment system for the GEM Premier 7000. portable blood analyzer
- Lot or code
- GEM Premier 7000: Part number: 00000015279 UDI: 0195226000924 Units distributed in US: GEM Premier 7000 Serial Nos. 26012062, 26012065, 26012069 and 26012070. Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634
- FDA recall number
- Z-2822-2026
- Quantity
- 4 US
Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 portable blood analyzer system
- Lot or code
- GEM Premier 5000: Part number: 00024019255 UDI: 08426950870227 Units distributed in Hong Kong: GEM Premier 5000 Serial Nos.: 20106073, 20106039, 19052933, 20116415, 20064792, 22062065, 22062067, 19103490, 22092716, 22102881, 18112339, 19012518, 20126582, 20075256, 20064794, 20106036, 24016057, 21088758, 21088759, 18061910, 24016081, 24016108, 24108036, 24108044, 24067370, 24067371 and 24077468. Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634
- FDA recall number
- Z-2823-2026
- Quantity
- 27
Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 portable blood analyzer system
- Lot or code
- GEM Premier 5000: Part number: 00024019255 UDI: 08426950870227 Units distributed in United Kingdom: GEM Premier 5000 Serial Nos.: 25110547, 25100462, 25110522, 25100472, 25101897, 25101920, 25101931, 25122029, 26012032, 25122030, 26012033, 26012043, 25110523, 25110536, 25120550, 25069638, 23101109, 24011137, 25081799, 25081797, 25091839, 25091840, 25091864, 25091870, 25091868, 25091873, 25091880, 25091876, 25101886, 25101889, 25101887, 25101891, 25101907, 25101904, 25101912, 25101905, 25101917, 25101921, 25101926, 25101929, 25101934, 25101927, 25111968, 25121986, 25121985, 25121981, 25121984, 25121994, 25121995, 25121982, 25121993, 25121991, 25121999, 25121996, 16030255, 17060782, 17060784, 17060785, 17060789, 17060790, 17060797, 17111126, 17121258, 18011414, 20054515, 20054516 and 22082271. Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634
- FDA recall number
- Z-2824-2026
- Quantity
- 67
Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Portable blood analyzer system
- Lot or code
- GEM Premier 5000: Part number: 00024019255 UDI: 08426950870227 Units distributed in Spain: GEM Premier 5000 Serial Nos.: 25028719, 18021546, 18021549, 26012036, 18051850, 23054764, 21068342, 25080219, 26010781, 25122012, 25041639, 25120558, 25079944, 25079943, 25028719, 25100485, 25079947, 25100483, 25100484, 21057870 and 26010710. Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634
- FDA recall number
- Z-2825-2026
- Quantity
- 21
Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Portable blood analyzer system
- Lot or code
- GEM Premier 5000: Part number: 00024019255 UDI: 08426950870227 Units distributed in Argentina: GEM Premier 5000 Serial Nos.: 25100466, 24097933 and 23064897. Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634
- FDA recall number
- Z-2826-2026
- Quantity
- 3
Distribution (FDA): US Nationwide distribution.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Instrumentation Laboratory
- Title