FDA recall 98962: TELEFLEX LLC: Brand Name: Pilling Wecksorb Product Name: Pilling Wecksorb Neurosponges

FDA recall 98962 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Due to out-of-specification, or potential out-of-specification, levels of endotoxins
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

194 more codes are in the record below.

  • Brand Name: Pilling Wecksorb Product Name: Pilling Wecksorb Neurosponges (Cotton) Pilling¿ Wecksorb Neurosponge Strips (Cotton) Model/Catalog Number: 200100 200101 200102 200103 200104 200105 200106 200108 200110 200112 200120 200121 200123 200131 Software Version: N/A Product Description: Wecksorb Neurosponges are made of an absorbent cotton material. Certain Neurosponge varieties are available with an x-ray detectable thread to assist with location within the surgical site. Some Neurosponges (patties) have a string for accountability during the post-surgical verification process. Wecksorb Neurosponges are for use in the protection of tissue during surgical procedures; tissues include the brain and other central nervous system tissues. Neuro sponges are used during cranial and spine surgery to moisten the dura or to cushion against bone edges. They are also used to protect brain tissue under the dura or elsewhere during incision. Component: N/A

    Lot or code
    Pilling¿ Wecksorb Neurosponges (Cotton), 1/2" x 1/2" X-Ray Detectable Catalog Number: 200100 UDI codes: (01)2 4026704 41748 7(17)260521(11)210521(10)21E1485 (01)2 4026704 41748 7(17)260811(11)210811(10)21H2304 (01)2 4026704 41748 7(17)260906(11)210906(10)21H2594 (01)2 4026704 41748 7(17)260906(11)210906(10)21H2595 (01)2 4026704 41748 7(17)261018(11)211018(10)21K3161 (01)2 4026704 41748 7(17)261130(11)211130(10)21L3593 (01)2 4026704 41748 7(17)270203(11)220203(10)22A0175 (01)2 4026704 41748 7(17)270316(11)220316(10)22C0661 (01)2 4026704 41748 7(17)270512(11)220512(10)22E1386 (01)2 4026704 41748 7(17)270915(11)220915(10)22J2777 (01)2 4026704 41748 7(17)271019(11)221019(10)22K3102 (01)2 4026704 41748 7(17)271027(11)221027(10)22K3269 (01)2 4026704 41748 7(17)271031(11)221031(10)22K3270 (01)2 4026704 41748 7(17)280124(11)230124(10)23A0102 (01)2 4026704 41748 7(17)280405(11)230405(10)23D0796 (01)2 4026704 41748 7(17)280426(11)230426(10)23D0900 (01)2 4026704 41748 7(17)280530(11)230530(10)23E1180 (01)2 4026704 41748 7(17)280809(11)230809(10)23H1834 (01)2 4026704 41748 7(17)290319(11)240319(10)24C0647 (01)2 4026704 41748 7(17)290522(11)240522(10)24E1133 (01)2 4026704 41748 7(17)290729(11)240729(10)24G1865 (01)2 4026704 41748 7(17)290731(11)240731(10)24G1864 (01)2 4026704 41748 7(17)291112(11)241112(10)24K2820 Lot/Batch Numbers: 21E1485 21H2304 21H2594 21H2595 21K3161 21L3593 22A0175 22C0661 22E1386 22J2777 22K3102 22K3269 22K3270 23A0102 23D0796 23D0900 23E1180 23H1834 24C0647…
    FDA recall number
    Z-2639-2026
    Quantity
    1488700

    The text above is the beginning of the record; 10,605 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The country of Panama.

Source

Issued by
U.S. Food and Drug Administration
Firm
TELEFLEX LLC
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Month: July 2026, Week 2026-W28. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Due to out-of-specification, or potential out-of-specification, levels of endotoxins”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 98962 on July 8, 2026.