What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050
- Lot or code
- UDI:00860009268616 Lot: 25G263
- FDA recall number
- Z-2275-2026
- Quantity
- 5904 units
Distribution (FDA): US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, KY, MI, MS, NC, NJ, NY, TN, TX, UT, VA.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Oculus Technologies of Mexico, S.A. de C.V.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.