What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.
- Lot or code
- Product Code LVP-SW-0005. UDI-DI: 00811505030122. Software Version 5.10.2
- FDA recall number
- Z-2223-2026
- Quantity
- 32 systems
Distribution (FDA): US Nationwide distribution in the states of CA, CO, FL, GA, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Fresenius Kabi USA, LLC
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.