What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Under extended clinical use conditions (e.g., long duration studies), moisture generated within the pneumatic circuit may accumulate inside the system's internal tubing or at the hydrophobic filter and waterproof pressure sensors. Accumulated moisture may affect airflow or pressure measurement performance if not properly removed.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
5 more codes are in the record below.
Solar GI System (Water Perfusion); Product Name (Catalog Number): Solar GI LGI H2O, 12p, Trolley system (G3-6), Solar GI HRM H2O, 24p, Compact Pole system (G3-7), Solar GI HRM H2O, 24p, Trolley system (G3-8), Solar GI HRM H2O, 36p, Trolley system (G3-12), Solar GI HRAM H2O, 24p, Trolley system (G3-14);
- Lot or code
- 1. Catalog Number: G3-6; UDI-DI: Primary UDI-DI: 08717775955047; Basic UDI-DI: 87177759500G3G8; All serial numbers; 2. Catalog Number: G3-7; UDI-DI: Primary UDI-DI: 08717775955054; Basic UDI-DI: 87177759500G3G8; All serial numbers; 3. Catalog Number: G3-8; UDI-DI: Primary UDI-DI: 08717775955061; Basic UDI-DI: 87177759500G3G8; All serial numbers; 4. Catalog Number: G3-12; UDI-DI: Primary UDI-DI: 08717775955078; Basic UDI-DI: 87177759500G3G8; All serial numbers; 5. Catalog Number: G3-14; UDI-DI: Primary UDI-DI: 08717775955085; Basic UDI-DI: 87177759500G3G8; All serial numbers;
- FDA recall number
- Z-2669-2026
- Quantity
- 372 units
Distribution (FDA): Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Canada, China, Costa Rica, Croatia, Cyprus, Czech Republic. Denmark, Egypt, France, Germany, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Lithuania, Malaysia, Namibia, Nepal, Netherlands. Oman, Osterreich, Pakistan, Poland, Portugal, Puerto Rico, Qatar, Russia, Russian Federation, Saudi Arabia, Serbia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Tunisia, Turkey, UAE, United Kingdom, and Vietnam.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Laborie Medical Technologies Corp
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.