What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
The V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face. The Impactor V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Arvis Hip & Knee 3.0 Instrument Set, REF: IN-28000 containing one of the following components: IN-10500, Impactor V-Block Assembly; IN-10600 , Tracker B; IN-16400, Tracker A; IN-16100, Tracker C; IN-28050, Tracker E
- Lot or code
- All Lots/serial Numbers; UDI: IN-28000: 00810180350020; IN-10500: (01)00810180350686; IN-10600 (01)00810180350693; IN-16400:(01)00810180350525; IN-16100: (01)00810180350563; IN-28050: 00810180350051
- FDA recall number
- Z-2292-2026
- Quantity
- 69 units
Arvis Hip & Knee 2.0 Instrument Set, REF: IN-21200 containing one of the following components: IN-10500, Impactor V-Block Assembly; IN-10600 , Tracker B; IN-16400, Tracker A; IN-16100, Tracker C
- Lot or code
- All Lots/serial Numbers; UDI: IN-28000: 00810180350020; IN-10500: (01)00810180350686; IN-10600 (01)00810180350693; IN-16400:(01)00810180350525; IN-16100: (01)00810180350563;
- FDA recall number
- Z-2293-2026
- Quantity
- 124 kits
Distribution (FDA): Worldwide - US Nationwide distribution in the states of MS, CA, IA, IN, MI, CO, MT, FL, IL, NY, AL, WI, NE, AZ, MN, VT, NH, MD, RI, MA, NJ, PA, KS, WA, ID, OR, TN, VA, PR and the countries of Australia, France, Germany, Italy.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Kico Knee Innovation Company
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.