What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Affected kits contain Medline Namic RA Syringes, which are recalled by their manufacturer due to risk of syringe rotating adapter unwinding during use, which may result in a loose connection or full disconnection between syringe and manifold. If unwinding occurs, there is a potential for biohazard exposure, blood loss, infection. or air embolism.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Custom Convenience Kit containing recalled Medline Namic Angiographic Control Syringes with Rotating Adapter (Namic RA Syringes) Angio Pack Part numbers AMS6908E, AMS6908F
- Lot or code
- Pack AMS6908E UDI-DI B098AMS6908E0 Lot 214875 Pack AMS6908F UDI-DI B098AMS6908F0 Lots 215436, 225001, 226706, 234088
- FDA recall number
- Z-2066-2026
- Quantity
- 150 kits (US only)
Distribution (FDA): US Nationwide distribution in the state of Arizona.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Windstone Medical Packaging, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.