What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Potential for mislabeled syringe produced by the i.v.STATION device.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Rolls of label stock used with the Omnicell i.v.Station. 1. Part Number 258920028, LABEL, STERILE, IVSTATION SYRINGE (40MMX40MM). 2. Part Number 258900029, LABEL, STERILE, IVSTATION SYRINGE (20MMX40MM). Model/Catalog Number: IVS-RBT-001. i.v.STATION is an automation device used for preparation and compounding of IV medications, which also applies labels to compounded preparations for identification and traceability.
- Lot or code
- 1. Part Number 258920028, Lot Number: QN 200023473. 2. Part Number 258900029, Lot Number: QN 200023474.
- FDA recall number
- Z-2047-2026
- Quantity
- 220 units
Distribution (FDA): US Distribution in Alabama, Maryland and Pennsylvania.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Omnicell, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.