Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 98695: HF Acquisition Co LLC: Brand Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE and 1 more

FDA recall 98695 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Due to incorrect incubation process being performed on product.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

  • Brand Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 52 TESTS Product Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 52 TESTS Model/Catalog Number: 1023780 Software Version: N/A Product Description: The SporeTraq Mail-In Spore Testing Service includes 52 bacterial Spore Strips provided by Mesa Labs. The Spore Strip (single species G. stearothermophilus ATCC#7953 or dual species G. stearothermophilus ATCC #7953 and B. atrophaeus ATCC #9372) is intended for use in testing the efficacy of sterilizer systems used in healthcare facilities. The Spore Strips are compatible with Steam sterilization, Chemiclave¿ sterilization, Ethylene Oxide sterilization, or Dry Heat sterilization processes. The test strip is placed in the customer s sterilizer during their sterilization cycle and upon completion the sterilized test strip along with the control strip is returned to Mesa Labs for lab testing in the return envelope. Spore test results are digitally stored in HealthFirst s OnTraq smart automation tracking and reporting system for easy and secure access to a paperless archive from any device with internet access. Component: N/A

    Lot or code
    Lot Code: Please see attached spreadsheet listing products that were not incubated per the following requirements : a. Spore strips used in steam sterilizers are incubated for 24 hours b. Spore strips used in chemiclave sterilizers are incubated for 72 hours c. Spore strips used in dry heat or ethylene oxide sterilizers are incubated for 7 days As noted in the recall scope question, HealthFirst's Field Correction scope will be expanded to include notifications to all customers in possession of 1023780 SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 52 TESTS,1023770 SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS or other products serviced by Mesa Labs. Product/Item Number: 1023780 UDI-DI code: 00850065007496 Lot Numbers: DSS-122 DSS-125 DSS-128 DSS-131 DSS-134
    FDA recall number
    Z-2331-2026
    Quantity
    2210
  • Brand Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS Product Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS Model/Catalog Number: 1023770 Software Version: N/A Product Description: The SporeTraq Mail-In Spore Testing Service includes 12 bacterial Spore Strips provided by Mesa Labs. The Spore Strip (single species G. stearothermophilus ATCC#7953 or dual species G. stearothermophilus ATCC #7953 and B. atrophaeus ATCC #9372) is intended for use in testing the efficacy of sterilizer systems used in healthcare facilities. The Spore Strips are compatible with Steam sterilization, Chemiclave¿ sterilization, Ethylene Oxide sterilization, or Dry Heat sterilization processes. The test strip is placed in the customer s sterilizer during their sterilization cycle and upon completion the sterilized test strip along with the control strip is returned to Mesa Labs for lab testing in the return envelope. Spore test results are digitally stored in HealthFirst s OnTraq smart automation tracking and reporting system for easy and secure access to a paperless archive from any device with internet access. Component: N/A

    Lot or code
    Lot Code: Please see attached spreadsheet listing products that were not incubated per the following requirements : a. Spore strips used in steam sterilizers are incubated for 24 hours b. Spore strips used in chemiclave sterilizers are incubated for 72 hours c. Spore strips used in dry heat or ethylene oxide sterilizers are incubated for 7 days As noted in the previous recall scope question, HealthFirst's Field Correction scope will be expanded to include notifications to all customers in possession of 1023780 SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 52 TESTS,1023770 SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS or other products serviced by Mesa Labs.
    FDA recall number
    Z-2332-2026
    Quantity
    0

Distribution (FDA): U.S. Nationwide distribution including in the states of AR, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, and WI.

Source

Issued by
U.S. Food and Drug Administration
Firm
HF Acquisition Co LLC
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: HF Acquisition Co LLC, Month: June 2026, Week 2026-W25. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Due to incorrect incubation process being performed on product.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 98695 on June 17, 2026.