What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Potential for spontaneous closures of Hamilton HxRunControl software during execution of Autolys protocols after deployment of the PCT software for AutoLys Hamilton STAR.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Protocol Configuration Tool (PCT) software v1.3.0 for Hamilton AutoLys STAR used within FoundationOneCDx (F1CDx) installed on Research Triangle Park laboratory Equipment IDs 01342, 01591, 01989, 01990, 01994, 01995.
- Lot or code
- Protocol Configuration Tool (PCT) software v1.3.0 for Hamilton AutoLys STAR used within FoundationOneCDx (F1CDx)
- FDA recall number
- Z-1937-2026
- Quantity
- 10 units
Distribution (FDA): US distribution to North Carolina and Massachusetts.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Foundation Medicine, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.