What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
FDA classification: Class III. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Heraeus, PALAMIX uno. Material Number: 66057893.
- Lot or code
- Material Number: 66057893; UDI-DI: 04260102133587; Batch: all batches with numbers below 223060;
- FDA recall number
- Z-1997-2026
- Quantity
- 51,771 units
Heraeus, PALAMIX duo. Material Number: 66057897.
- Lot or code
- Material Number: 66057897; UDI-DI: 04260102133617; Batch: all batches with numbers below 520215;
- FDA recall number
- Z-1998-2026
- Quantity
- 1,125 units
Distribution (FDA): Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Korea, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, Uruguay, Utd.Arab Emir.;
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Heraeus Medical GmbH (Dental Division)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.