What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of Previous Surview Exists Select Previous Surview? <Yes> or <No> should display but does not appear. 3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Philips Spectral CT on Rails. Model Number: 728334.
- Lot or code
- Model Number: 728334. UDI: (01)00884838103627(21)1005, (01)00884838103627(21)1006, (01)00884838103627(21)1007. Software Version Number: 5.1.X. Serial Numbers: 1005, 1006, 1007.
- FDA recall number
- Z-1825-2026
- Quantity
- 3 systems
Distribution (FDA): Worldwide - US Nationwide distribution in the states of Minnesota and the countries of France and Netherlands.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- PHILIPS MEDICAL SYSTEMS
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.