Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 98542: ARJOHUNTLEIGH POLSKA Sp. z.o.o.: Arjo Tenor mobile passive patient lift

FDA recall 98542 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

An internal component of the actuators in a certain production batch may not meet the required mechanical strength. This could result, in the worst case scenario, in a sudden loss of the actuator's ability to hold the load, leading to rapid and uncontrolled downward movement of the lifting arm.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

115 more codes are in the record below.

  • Arjo Tenor mobile passive patient lift, Model Numbers KHA1000 US and KHA1010 US

    Lot or code
    UDI-DI: 05055982787758 and 05055982787710 Serial Number: P0911221 P0909831 P0925191 P0907166 P0901255 P0906820 P0925431 P0912820 P0924568 P0910471 P0903999 P0914725 P0909539 P0910797 P0896659 P0911350 P0925190 P0914588 P0904321 P0905040 P0924559 P0913229 P0909537 P0923376 P0923377 P0911222 P0899348 P0909201 P0900805 P0912570 P0903592 P0912571 P0912573 P0914003 P0900806 P0910470 P0907737 P0907738 P0914004 P0908250 P0908429 P0914122 P0916220 P0912343 P0912821 P0924239 P0925203 P0925433 P0903476 P0912342 P0913326 P0904320 P0905039 P0897003 P0915180 P0898734 P0907165 P0907736 P0897841 P0903997 P0925432 P0896658 P0915498 P0914724 P0915163 P0913227 P0914118 P0908428 P0908768 P0914121 P0912572 P0915500 P0909200 P0911602 P0909202 P0912341 P0906821 P0899349 P0905354 P0905355 P0905356 P0905980 P0914726 P0908251 P0916221 P0910795 P0907167 P0907507 P0914589 P0897247 P0914587 P0903591 P0911601 P0903477 P0913228 P0910796 P0908430 P0903998 P0904319 P0904739 P0904740 P0899214 P0923462 P0905041 P0911603 P0907505 P0907506 P0898735 P0901254 P0912822 P0909830 P0910469 P0914586 P0899215 P0900415 P0897002 P0897248 P0914002 P0897842
    FDA recall number
    Z-1811-2026
    Quantity
    119 units

Distribution (FDA): US Nationwide distribution.

Source

Issued by
U.S. Food and Drug Administration
Firm
ARJOHUNTLEIGH POLSKA Sp. z.o.o.
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: ARJOHUNTLEIGH POLSKA Sp. z.o.o., Month: April 2026, Week 2026-W17. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “An internal component of the actuators in a certain production batch may not meet the required mechanical strength. This could result, in the worst case scenario, in a sudden loss of the actuator's ability to hold the load, leading to rapid and uncontrolled downward movement of the lifting arm.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 98542 on April 22, 2026.

Other notices of ARJOHUNTLEIGH POLSKA Sp. z.o.o.