Why it was recalled
In certain situations, following a system reboot, one or more modules may have become unreachable via the remote connectivity interface. When this occurred, there was potential for delays in plate retrieval from the incubator while connectivity was restored. This issue was intermittent and did not affect test results, performance of the assay, or sample integrity.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
BD Kiestra" ReadA; Catalog No.: 446948.
- Lot or code
- Catalog No.: 446948; UDI-DI: 00382904469481; Serial Numbers: All;
- FDA recall number
- Z-1658-2026
- Quantity
- 82 units (7 US, 65 OUS)
Distribution (FDA): Domestic: KS, MD, MN, NJ, TX, VA; International: Australia, Austria, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, Italy, Kuwait, Mexico, Morocco, Netherlands, Norway, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, United Kingdom;