What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
If heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by placing pad under the body, then the high limit device is unable to sense the higher temperature so power is not terminated. May cause a burning smell prior to melting/burning potentially causing burns to persons and property.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Handy Solutions Neck & Shoulders Heating Pad, Number: 25607, Manufactured by Ninbo Unico Products Co., Ltd
- Lot or code
- UPC: 0-24291-25607-6, UDI-DI: 00024291256076
- FDA recall number
- Z-1588-2026
- Quantity
- 28,457
Distribution (FDA): US Nationwide distribution in the states of CA, MS, FL, IN, LA, VA, TX.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Navajo Manufacturing Company
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.