What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.
FDA classification: Class III. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
22 more codes are in the record below.
Battery Charging Station; Model: 0998-00-0802;
- Lot or code
- Model: 0998-00-0802; UDI: 10607567111964; Serial Numbers: 200019347EAI, 200019847EAI, 200019447EAI, 210018625EAI, 210017925EAI, 210017525EAI, 210018425EAI, 210020144EAI, 210020944EAI, 210022544EAI, 210020244EAI, 210022144EAI, 210020044EAI, 210020544EAI, 210021344EAI, 210023244EAI, 210018025EAI, 210019025EAI, 210021144EAI, 210021044EAI, 210022344EAI, 210019225EAI, 210019125EAI, 210018525EAI, 210022844EAI, 200017647EAI, 210019725EAI, 210022744EAI;
- FDA recall number
- Z-1587-2026
- Quantity
- 28 units (20 US, 8 OUS)
Distribution (FDA): Worldwide - US Nationwide distribution in the states of FL, DC, GA, IL, IN, MD, MI, NC, TX, WA, WI and the countries of Australia, Hong Kong, Israel, Japan, Norway.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Datascope Corp.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.