What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Fabric may experience premature delamination when using included laundering instructions. Identifying delaminated gown or drape in a surgical setting may result in brief delay in patient care. If delaminated product is used and surgical strikethrough occurs, there may be risk of infection.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Medline Medcrest Surgical Gowns: MDTGXC4J5XL DBQ-GOWN,SURG,XALT,L4, CRTCL CVR, GRN, 5 MDTGXC4JL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,L MDTGXC4JXL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,XL MDTGXC4JXXL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,XX MDTGXP4JL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,L MDTGXP4JXL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XL MDTGXP4JXXL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XX MDTL1002862EF DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XL MDTZ1000893AJ DBD-GOWN,SRG,PNL CVR, GREEN,SNPS,XL
- Lot or code
- UDI-DI (each/case) 10197344014002 20197344014009 10888277189188 40888277189189 10884389189191 40884389189192 10884389193297 40884389193298 10888277189218 40888277189219 10884389189221 40884389189222 10193489040111 20193489040118 10884389189221 40884389189222 10193489021820 20193489021827 All lots affected
- FDA recall number
- Z-1563-2026
- Quantity
- 104,004
Medline Surgical Drapes: MDTBTCS6070GN DBD-DRAPE,COVER,TABLE,RESIS/XALT GRN,60X MDTBTCS6090GN DBD-DRAPE,COVER,TABLE,RESIS/XALT GRN,60X MDTDXUNDBGN DBD-DRAPE,UNDERBUTTOCK,28 X 50.5,XALT GR MDTMAYOXGN DBD-DRAPE,COVER,MAYO STAND,RESISTAT/XALT MDTZ1003656GN DBD-POCKET SHEET,RESISTAT/XALT
- Lot or code
- UDI-DI (each/case) 10193489036961 20193489036968 10193489036978 20193489036975 10193489040098 20193489040095 10193489036954 20193489036951 10193489036985 20193489036982 All lots affected
- FDA recall number
- Z-1564-2026
- Quantity
- 23,568
Distribution (FDA): Domestic distribution nationwide US. International distribution to Panama, United Arab Emirates, Virgin Islands, Saudi Arabia, India, Chile, and Canada.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Medline Industries, LP (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.