What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
MICS3 Angled Sagittal Saw Attachment; Part Number: 210490
- Lot or code
- Part Number: 210490; UDI: (01)07613327523874(10)TU6563205RP; Lot numbers: ALL;
- FDA recall number
- Z-1572-2026
- Quantity
- 1098 units
Distribution (FDA): Domestic: Nationwide Distribution; International: Australia, Germany, Belgium, France, Spain UK, New Zealand;
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Howmedica Osteonics Corp. (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.