Why it was recalled
It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
IBA Proton Therapy System - PROTEUS 235
- Lot or code
- PROTEUS 235; UDI-DI: 05404013801138; Serial Numbers: SBF103 (JP), SBF124 (EU), SBF133 (US), SBF140 (EU)
- FDA recall number
- Z-1497-2026
- Quantity
- 4 units (1 US, 3 OUS)
Distribution (FDA): Worldwide - US Nationwide distribution in the state of FL and the countries of Italy and Japan.