Why it was recalled

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

124 more codes are in the record below.

  • Medline Kits containing Tego Connectors: Medline IR Pack, SKU DYNJ67205C Medline NO CATHETER DIALYSIS INSERTION, SKU CVI5200

    Lot or code
    IR PACK UDI-DI (ea) 10195327621780 UDI-DI (case) 40195327621781 Kit lots 24BBP627 24DBO029 24FBI148 24JBA858 24KBD925 24LBG978 25ABQ002 25EBV162; NO CATHETER DIALYSIS INSERTION UDI-DI (ea) 10195327675219 UDI-DI (case) 40195327675210 Kit Lots 24EBM915 24GBC287
    FDA recall number
    Z-1463-2026
    Quantity
    516
  • Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746

    Lot or code
    UDI-DI (ea) 10653160373582 UDI-DI (case) 00653160373585 Kit lots 2024051380 2024052080 2024052180 2024052280 2024071580 2025061790 2025072990
    FDA recall number
    Z-1464-2026
    Quantity
    2020
  • Convenience kits used for dialysis maintenance Medline ADD A CATH DIALYSIS KIT SKU ECVC8415A Centurion CENTRAL LINE INSERTION TRAY SKU DT19810 Centurion DIALYSIS BUNDLE WITH NO CATHETTER SKU CVI4310A Medline DIALYSIS CAP CHANGE KIT SKU DYNDC2425 Medline DIALYSIS CHANGE KIT SKU EBSI1453A Centurion HEMODIALYSIS ACCESS KIT SKU DT12970 Medline LARGE BORE, DIALYSIS, ENHANCED SECUREMENT SKU EBSI1741

    Lot or code
    Medline ADD A CATH DIALYSIS KIT SKU ECVC8415A UDI-DI (ea) 10653160383628 UDI-DI (case) 00653160383621 Kit lots 2024121390 2025033190 Centurion CENTRAL LINE INSERTION TRAY SKU DT19810 UDI-DI (ea) 10653160368755 UDI-DI (case) 00653160368758 Kit lot 2025041090 Centurion DIALYSIS BUNDLE WITH NO CATHETTER SKU CVI4310A UDI-DI (ea) 10653160374978 UDI-DI (case) 00653160374971 Kit lots 2024051490 2025021290 2025051390 2025091090 Medline DIALYSIS CAP CHANGE KIT SKU DYNDC2425 UDI-DI (ea) 10653160375845 UDI-DI (case) 00653160375848 Kit lots 2024060690 2024101690 2024103190 2025010890 2025040190 Medline DIALYSIS CHANGE KIT SKU EBSI1453A UDI-DI (ea) 10653160383475 UDI-DI (case) 00653160383478 Kit lots 2024050790 2024051390 2024082290 2024111990 2025013090 2025032090 2025050190 2025062790 Centurion HEMODIALYSIS ACCESS KIT SKU DT12970 UDI-DI (ea) 10653160368922 UDI-DI (case) 00653160368925 Kit lots 2025040401 2025081401 Medline LARGE BORE, DIALYSIS, ENHANCED SECUREMENT SKU EBSI1741 UDI-DI (ea) 10653160374527 UDI-DI (case) 00653160374520 Kit lots 2024061780 2024070180 2024081280 2024111290 2025012490 2025032690 2025051390 2025061790 2025071490 2025072190
    FDA recall number
    Z-1465-2026
    Quantity
    14,525
  • Medline kits containing Tego Connectors Medline APHERESIS SUBQPORT ACCESS KIT SKU DYNDA2705 Medline CENTRAL LINE INSERTION SKU DYNJ63347A

    Lot or code
    SKU DYNDA2705 UDI-DI (ea) 10193489468632 UDI-DI (case) 40193489468633 Kit lots 24DBL787 24HBR630 25ABH699 25DBV209 25GBQ405 SKU DYNJ63347A UDI-DI (ea) 10193489278064 UDI-DI (case) 40193489278065 Kit lots 24AMC874 24AME139 24AMH820 24BMF097 24CMF691 24GMB769 24GMK036 24IMC124 24LMA110 25AMJ634 25GMD386 25IMB510 25JMI708
    FDA recall number
    Z-1466-2026
    Quantity
    1212
  • Convenience kits containing Tego Connectors: Medline DIALYSIS DRESSING CHANGE SKU EBSI1498 Medline DIALYSIS DRESSING CHANGE KIT SKU EBSI1107 EBSI1596 EBSI1781 EBSI1812 Centurion DIALYSIS/PHERESUS CVC DRSG CHANGE KIT SKU DT22630 Medline DIALYSIS/PHERESIS DRESSING CHANGE KIT SKU EBSI1470 Centurion VASCATH DRESSING CHANGE TRAY SKU DT22845

    Lot or code
    Medline DIALYSIS DRESSING CHANGE SKU EBSI1498 UDI-DI (ea) 10193489111538 UDI-DI (case) 20193489111535 Kit lots 2024052090 2024072690 2024100390 2024121290 2025020490 2025041590 2025062790 2025090890 Medline DIALYSIS DRESSING CHANGE KIT SKU EBSI1107 UDI-DI (ea) 10889942723508 UDI-DI (case) 40889942723509 Kit lots 2024051690 2024073190 2024092490 2024121190 2025020790 2025032790 2025052990 EBSI1596 UDI-DI (ea) 10653160357162 UDI-DI (case) 00653160357165 Kit lots 2024070190 2024082390 2024102390 2024121790 2025031090 2025040190 2025052790 2025072890 EBSI1781 UDI-DI (ea) 10653160379638 UDI-DI (case) 00653160379631 Kit lots 2024100490 2024123090 EBSI1812 UDI-DI (ea) 10653160386735 UDI-DI (case) 00653160386738 Kit lots 2025032190 2025052990 Centurion DIALYSIS/PHERESUS CVC DRSG CHANGE KIT SKU DT22630 UDI-DI (ea) 10653160345343 UDI-DI (case) 00653160345346 Kit lots 2024050680 2024072980 2024090380 2024111290 2024121190 2025032890 2025061790 2025072390 Medline DIALYSIS/PHERESIS DRESSING CHANGE KIT SKU EBSI1470 UDI-DI (ea) 10193489199307 UDI-DI (case) 20193489199304 Kit lots 2024052080 2024070880 2024110490 2025010290 2025022090 2025030390 2025040990 2025072990 Centurion VASCATH DRESSING CHANGE TRAY SKU DT22845 UDI-DI (ea) 10653160355151 UDI-DI (case) 00653160355154 Kit lots 2024060790 2024080790 2024100990 2024112590 2025042490 2025082090
    FDA recall number
    Z-1467-2026
    Quantity
    31,848
  • Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167

    Lot or code
    UDI-DI (ea) 10195327015954 UDI-DI (case) 40195327015955 Kit lots 24IME953 25CMA708
    FDA recall number
    Z-1468-2026
    Quantity
    150

Distribution (FDA): US Nationwide distribution.

Source

Issued by
U.S. Food and Drug Administration
Firm
Medline Industries, LP (other notices of this firm)
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Medline Industries, LP, Month: March 2026, Week 2026-W10. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 98333 on March 4, 2026.

Other notices of Medline Industries, LP