What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.
- Lot or code
- Model No. 7303, UDI-DI: 00840095607575, Lot No. 25XXX.
- FDA recall number
- Z-1370-2026
- Quantity
- 1,578 units
Distribution (FDA): International distribution to the countries of Russia, Bulgaria, Kazakhstan, Ukraine, Pakistan, Belarus.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Medica Corporation
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.