What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Reports of device failure, including anchor breakage, anchor bending, anchor pull out, suture breakage, and secondary packaging breakage.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
11 more codes are in the record below.
Fixone Hybrid Anchor
- Lot or code
- Model No. KAPN-47501nc, KAPN-55001c, KAPN-55001na, KAPN-55001nc; All UDI Codes; All Lots.
- FDA recall number
- Z-1166-2026
- Quantity
- 233 units
Fixone Biocomposite Anchor
- Lot or code
- Model No. BAB-55003ch, BAN-55003b, SAB-30002e; All UDI Codes; All Lots.
- FDA recall number
- Z-1167-2026
- Quantity
- 1,088 units
Genesis Screw-In Anchor
- Lot or code
- Model No. 10B45003, 10B55003, 10B65003; All UDI Codes; All Lots.
- FDA recall number
- Z-1168-2026
- Quantity
- 600 units
Genesis Dual Thread Screw-In Anchor
- Lot or code
- Model No. 10N47502, 10N55002; All UDI Codes; All Lots.
- FDA recall number
- Z-1169-2026
- Quantity
- 400 units
Genesis Knotless Anchor
- Lot or code
- Model No. 10KPN4751, 10KPN5501; All UDI Codes; All Lots.
- FDA recall number
- Z-1170-2026
- Quantity
- 1,163 units
Genesis Push-In Suture Anchor
- Lot or code
- Model No. 10S30002; All UDI Codes; All Lots.
- FDA recall number
- Z-1171-2026
- Quantity
- 200 units
AlternatiV+ Max Knotless Anchor
- Lot or code
- Model No. 20KPN4751, 20KPN5501; All UDI Codes; All Lots.
- FDA recall number
- Z-1172-2026
- Quantity
- 2,569 units
AlternatiV+ Screw-In Anchor
- Lot or code
- Model No. 20BN4753, 20BN5503; All UDI Codes; All Lots.
- FDA recall number
- Z-1173-2026
- Quantity
- 416 units
Distribution (FDA): US Nationwide distribution in the states of CA & PR.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Aju Pharm Co., Ltd.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.