Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 98226: Medline Industries, LP: Medline medical procedure convenience kits labeled as: NERVE BLOCK TRAY 2 and 14 more

FDA recall 98226 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

385 more codes are in the record below.

  • Medline medical procedure convenience kits labeled as: NERVE BLOCK TRAY 2, Kit SKU DYNJRA2713

    Lot or code
    UDI/DI 10198459565861 (EA) 40198459565862 (CS), Lot number 25KMD285
    FDA recall number
    Z-1186-2026
    Quantity
    0
  • Medline medical procedure convenience kits labeled as: 1) CATH LAB PACEMAKER PACK-LF, Kit SKU DYNJ64837C; 2) LOOP RECOVERY PACK, Kit SKU DYNJT7034;

    Lot or code
    Medline Kit SKU DYNJ64837C: UDI/DI 10198459139482 (EA) 40198459139483 (CS), Lot Number 25KMJ193; Medline Kit SKU DYNJT7034: UDI/DI 10198459592591 (EA) 40198459592592 (CS), Lot Number 25KMI766;.
    FDA recall number
    Z-1187-2026
    Quantity
    1,928 total
  • Medline medical procedure convenience kits labeled as: 1) PACER / CATH LAB PACK-LF, Kit SKU DYNJ0972604A; 2) PACEMAKER/ICD PACK, Kit SKU DYNJ33822L; 3) VENOUS ACCESS PACK, Kit SKU DYNJ45640B; 4) PACEMAKER SET UP, Kit SKU DYNJ63377D; 5) PACK PACEMAKER CUSTOM,9909178, Kit SKU DYNJ65011D; 6) HEART BASIC OPEN PACK, Kit SKU DYNJ69143C; 7) SCC VEIN PACK, Kit SKU DYNJ69313; 8) CV PACK, Kit SKU DYNJ86007B.

    Lot or code
    Medline Kit SKU DYNJ0972604A: UDI/DI 10889942403905 (EA) 40889942403906 (CS), Lot Number 25KMD144; Medline Kit SKU DYNJ33822L: UDI/DI 10198459385582 (EA) 40198459385583 (CS), Lot Number 25KMD419; Medline Kit SKU DYNJ45640B: UDI/DI 10193489494983 (EA) 40193489494984 (CS), Lot Number 25KMD793; Medline Kit SKU DYNJ63377D: UDI/DI 10198459514579 (EA) 40198459514570 (CS), Lot Number 25KMI988; Medline Kit SKU DYNJ65011D: UDI/DI 10198459017018 (EA) 40198459017019 (CS), Lot Number 25KMH161; Medline Kit SKU DYNJ69143C: UDI/DI 10198459478505 (EA) 40198459478506 (CS), Lot Number 25LMB211; Medline Kit SKU DYNJ69313: UDI/DI 10193489931822 (EA) 40193489931823 (CS), Lot Number 25KMJ489; Medline Kit SKU DYNJ86007B: UDI/DI 10198459260452 (EA) 40198459260453 (CS), Lot Number 25KDB551.
    FDA recall number
    Z-1188-2026
    Quantity
    1,928 total
  • Medline medical procedure convenience kits labeled as: 1) BREAST-HERNIA-PORT CDS-LF, Kit SKU CDS984853L; 2) BREAST BIOPSY PACK, Kit SKU DYNJ45641B; 3) DERMATOLOGY PACK, Kit SKU DYNJT3658A.

    Lot or code
    Medline Kit SKU CDS984853L: UDI/DI 10198459293535 (EA) 40198459293536 (CS), Lot Number 25KMI898; Medline Kit SKU DYNJ45641B: UDI/DI 10193489494921 (EA) 40193489494922 (CS), Lot Number 25KMH748; Medline Kit SKU DYNJT3658A: UDI/DI 10198459354007 (EA) 40198459354008 (CS), Lot Number 25KME009.
    FDA recall number
    Z-1189-2026
    Quantity
    1,928 total
  • Medline medical procedure convenience kits labeled as: 1) MIDDLE EAR CDS, Kit SKU CDS984347O; 2) HEAD AND NECK CDS-LF, Kit SKU CDS984857N; 3) NECK PACK-LF, Kit SKU DYNJ0966130K; 4) NECKLINE PACK-LF, Kit SKU DYNJ51440B; 5) THYROID PACK, Kit SKU DYNJ58742J; 6) ENT I, Kit SKU DYNJ63781B; 7) NECK PACK, Kit SKU DYNJ82751B; 8) NOSE PACK, Kit SKU DYNJ82753C; 9) WMC COCHLEAR IMPLANT ADD ON, Kit SKU DYNJ907206G.

    Lot or code
    Medline Kit SKU CDS984347O: UDI/DI 10198459311451 (EA) 40198459311452 (CS), Lot Number 25KMD733; Medline Kit SKU CDS984347O: UDI/DI 10198459311451 (EA) 40198459311452 (CS), Lot Number 25KMC147; Medline Kit SKU CDS984857N: UDI/DI 10198459359705 (EA) 40198459359706 (CS), Lot Number 25KMJ151; Medline Kit SKU DYNJ0966130K: UDI/DI 10198459442117 (EA) 40198459442118 (CS), Lot Number 25LMB220; Medline Kit SKU DYNJ0966130K: UDI/DI 10198459442117 (EA) 40198459442118 (CS), Lot Number 25KME804; Medline Kit SKU DYNJ51440B: UDI/DI 10198459004230 (EA) 40198459004231 (CS), Lot Number 25KMI445; Medline Kit SKU DYNJ58742J: UDI/DI 10198459254710 (EA) 40198459254711 (CS), Lot Number 25KMC534; Medline Kit SKU DYNJ63781B: UDI/DI 10198459549595 (EA) 40198459549596 (CS), Lot Number 25KMB410; Medline Kit SKU DYNJ82751B: UDI/DI 10198459039157 (EA) 40198459039158 (CS), Lot Number 25LMB068; Medline Kit SKU DYNJ82751B: UDI/DI 10198459039157 (EA) 40198459039158 (CS), Lot Number 25KMI309; Medline Kit SKU DYNJ82753C: UDI/DI 10198459446887 (EA) 40198459446888 (CS), Lot Number 25KMC126; Medline Kit SKU DYNJ907206G: UDI/DI 10198459368028 (EA) 40198459368029 (CS), Lot Number 25KBL336.
    FDA recall number
    Z-1190-2026
    Quantity
    40 units
  • Medline medical procedure convenience kits labeled as: 1) SMJ LAP CHOLE PACK-LF, Kit SKU DYNJ19974P; 2) DEPAUL ABDOMINAL ENDOSCOPY-LF, Kit SKU DYNJ20027K; 3) LAPAROSCOPY PACK-LF, Kit SKU DYNJ20468J; 4) SMJ LAPAROSCOPY PACK, Kit SKU DYNJ25763N; 5) SMJ LAPAROSCOPY PACK, Kit SKU DYNJ25763O; 6) BWNBORO LAP CHOLE PACK-LF, Kit SKU DYNJ32836B; 7) MINOR SET UP PACK, Kit SKU DYNJ43272K; 8) PK, GEN LAPAROSCOPIC CHOLE, Kit SKU DYNJ43976C; 9) GV LAP CHOLE PACK, Kit SKU DYNJ46939Q; 10) LAP CHOLE PACK, Kit SKU DYNJ49688C; 11) GENERAL ROBOTIC LAP, Kit SKU DYNJ51347N; 12) LAP CHOLE PACK, Kit SKU DYNJ54081F; 13) PEDIATRIC MOSES, Kit SKU DYNJ58864B; 14) LAPAROTOMY PACK, Kit SKU DYNJ59313B; 15) CRMC LAP CHOLE DIAG LAP-LF, Kit SKU DYNJ61095F; 16) LAP GASTRIC PACK, Kit SKU DYNJ62065A; 17) MINOR PACK, Kit SKU DYNJ64174A; 18) GENERAL SURGERY PACK, Kit SKU DYNJ64258I; 19) GENERAL LAP SHARED-LF, Kit SKU DYNJ64786D; 20) CHOLE PACK, Kit SKU DYNJ67686B; 21) GEN SURG MINOR PACK, Kit SKU DYNJ69142F; 22) LAP CHOLE PACK, Kit SKU DYNJ81695B; 23) GENERAL LAPAROSCOPY, Kit SKU DYNJ900901K; 24) GEN MINOR, Kit SKU DYNJ900906L; 25) ROBOTIC GENERAL, Kit SKU DYNJ906861D; 26) RMC GENERAL LAPAROSCOPIC, Kit SKU DYNJ907087C; 27) LAP CHOLE LMC CDS, Kit SKU DYNJ911116A; 28) MINOR GENERAL PACK, Kit SKU DYNJT3884; 29) GENERAL LAPAROSCOPY PACK, Kit SKU DYNJT4504; 30) MINOR SURGERY PACK-LF, Kit SKU PHS109062D.

    Lot or code
    Medline Kit SKU DYNJ19974P: UDI/DI 10198459138195 (EA) 40198459138196 (CS), Lot Number 25KMB830; Medline Kit SKU DYNJ19974P: UDI/DI 10198459138195 (EA) 40198459138196 (CS), Lot Number 25KMB831; Medline Kit SKU DYNJ20027K: UDI/DI 10198459228957 (EA) 40198459228958 (CS), Lot Number 25KMG395; Medline Kit SKU DYNJ20027K: UDI/DI 10198459228957 (EA) 40198459228958 (CS), Lot Number 25KMH132; Medline Kit SKU DYNJ20468J: UDI/DI 10195327024338 (EA) 40195327024339 (CS), Lot Number 25KMF540; Medline Kit SKU DYNJ25763N: UDI/DI 10198459138218 (EA) 40198459138219 (CS), Lot Number 25KMH134; Medline Kit SKU DYNJ25763O: UDI/DI 10198459603655 (EA) 40198459603656 (CS), Lot Number 25KMI209; Medline Kit SKU DYNJ32836B: UDI/DI 10195327037369 (EA) 40195327037360 (CS), Lot Number 25KBF167; Medline Kit SKU DYNJ43272K: UDI/DI 10198459257131 (EA) 40198459257132 (CS), Lot Number 25KME892; Medline Kit SKU DYNJ43976C: UDI/DI 10193489806694 (EA) 40193489806695 (CS), Lot Number 25KMG962; Medline Kit SKU DYNJ46939Q: UDI/DI 10198459404047 (EA) 40198459404048 (CS), Lot Number 25JDA437; Medline Kit SKU DYNJ46939Q: UDI/DI 10198459404047 (EA) 40198459404048 (CS), Lot Number 25LDA059; Medline Kit SKU DYNJ49688C: UDI/DI 10198459155017 (EA) 40198459155018 (CS), Lot Number 25KMG180; Medline Kit SKU DYNJ51347N: UDI/DI 10198459386428 (EA) 40198459386429 (CS), Lot Number 25JBW506; Medline Kit SKU DYNJ54081F: UDI/DI 10198459172410 (EA) 40198459172411 (CS), Lot Number 25KMB295; Medline Kit SKU DYNJ58864B: UDI/DI…
    FDA recall number
    Z-1191-2026
    Quantity
    456 units

    The text above is the beginning of the record; 1,812 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Medline medical procedure convenience kits labeled as: 1) MEM MAXILLOFACIAL PACK-LF, Kit SKU DYNJ0660040R; 2) PLASTICS VCH, Kit SKU DYNJ39225K; 3) WT BREAST PACK, Kit SKU DYNJ46620I; 4) ENDO / AUG PACK, Kit SKU DYNJ59250I; 5) KMC PLASTIC/GEN PACK-LF, Kit SKU DYNJ65980J; 6) PLASTICS PACK, Kit SKU DYNJ68288A; 7) IMPLANT REMOVAL KIT, Kit SKU MNS13035; 8) PLASTICS PACK-LF, Kit SKU PHS41746D.

    Lot or code
    Medline Kit SKU DYNJ0660040R: UDI/DI 10889942220083 (EA) 40889942220084 (CS), Lot Number 25KMG652; Medline Kit SKU DYNJ39225K: UDI/DI 10198459102257 (EA) 40198459102258 (CS), Lot Number 25JMJ958; Medline Kit SKU DYNJ39225K: UDI/DI 10198459102257 (EA) 40198459102258 (CS), Lot Number 25KMJ566; Medline Kit SKU DYNJ46620I: UDI/DI 10198459404009 (EA) 40198459404000 (CS), Lot Number 25KBE927; Medline Kit SKU DYNJ59250I: UDI/DI 10198459359941 (EA) 40198459359942 (CS), Lot Number 25KMB767; Medline Kit SKU DYNJ65980J: UDI/DI 10198459423727 (EA) 40198459423728 (CS), Lot Number 25KMJ630; Medline Kit SKU DYNJ68288A: UDI/DI 10195327056995 (EA) 40195327056996 (CS), Lot Number 25KMF183; Medline Kit SKU MNS13035: UDI/DI 10653160990482 (EA) 40653160990483 (CS), Lot Number 25KME754; Medline Kit SKU PHS41746D: UDI/DI 10198459438820 (EA) 40198459438821 (CS), Lot Number 25KMD226.
    FDA recall number
    Z-1192-2026
    Quantity
    66 units
  • Medline medical procedure convenience kits labeled as: 1) DRAWER 1 AIRWAY, Kit SKU ACC010270A; 2) DRAWER 2, Kit SKU ACC010690; 3) DRAWER 3 SYR/DRS/MON/NG 1 OF 3, Kit SKU ACC011073; 4) HYPOSPADIUS PACK, Kit SKU DYNJ00282Q; 5) SJ NEURO PACK RFID-LF, Kit SKU DYNJ0101360K; 6) CENTRAL LINE PACK-LF, Kit SKU DYNJ0220136S; 7) SPINE PACK-LF, Kit SKU DYNJ0376635Q; 8) ORTHO SPINAL PACK-LF, Kit SKU DYNJ0460768T; 9) W.C. MINOR LAPAROTOMY PK-LF, Kit SKU DYNJ0509186J; 10) NEURO SHUNT PACK-LF, Kit SKU DYNJ0578815W; 11) ANCILLARY SPINE PACK-LF, Kit SKU DYNJ0610290T; 12) MINOR LAPAROTOMY PACK-LF, Kit SKU DYNJ0753436P; 13) KNEE SCOPE PACK-LF, Kit SKU DYNJ0890502R; 14) PEDIATRIC PACK-LF, Kit SKU DYNJ28212L; 15) ANTERIOR CERVICAL PACK, Kit SKU DYNJ31341P; 16) PEDIATRIC MINOR PROCEDURE PACK, Kit SKU DYNJ37615J; 17) LAPAROSCOPY PACK, Kit SKU DYNJ39215N; 18) MAJOR LAP VCH, Kit SKU DYNJ39217P; 19) NEURO PACK, Kit SKU DYNJ39220N; 20) CAS HAND PK, Kit SKU DYNJ39468G; 21) RFT ISC NOBLES MINOR PACK, Kit SKU DYNJ43954B; 22) PED OPEN HEART PACK, Kit SKU DYNJ45373G; 23) MAJOR GENERAL PACK-LF, Kit SKU DYNJ51436F; 24) MINOR PK-LF, Kit SKU DYNJ51437C; 25) VEIN ABLATION & EXCISION ASC, Kit SKU DYNJ52586C; 26) SHOULDER PACK, Kit SKU DYNJ56889B; 27) LAP CHOLE PACK, Kit SKU DYNJ57206F; 28) TRANSGENDER MASTECTOMY 2, Kit SKU DYNJ58257I; 29) SHOULDER ARTH PACK, Kit SKU DYNJ58410D; 30) MAJOR PACK, Kit SKU DYNJ58528J; 31) PACK BREAST, Kit SKU DYNJ60157F; 32) CH TRANSPHENOIDAL PACK, Kit SKU DYNJ61696I; 33) RR-SPINE…

    Lot or code
    Medline Kit SKU ACC010270A: UDI/DI 10889942219827 (EA) 40889942219828 (CS), Lot Number 25LDA153; Medline Kit SKU ACC010690: UDI/DI 10195327433307 (EA) 40195327433308 (CS), Lot Number 25LDA088; Medline Kit SKU ACC011073: UDI/DI 10198459569333 (EA) 40198459569334 (CS), Lot Number 25LDA072; Medline Kit SKU DYNJ00282Q: UDI/DI 10195327412029 (EA) 40195327412020 (CS), Lot Number 25KMC904; Medline Kit SKU DYNJ00282Q: UDI/DI 10195327412029 (EA) 40195327412020 (CS), Lot Number 25LMA071; Medline Kit SKU DYNJ0101360K: UDI/DI 10195327662066 (EA) 40195327662067 (CS), Lot Number 25KMC709; Medline Kit SKU DYNJ0220136S: UDI/DI 10198459566073 (EA) 40198459566074 (CS), Lot Number 25KMB403; Medline Kit SKU DYNJ0220136S: UDI/DI 10198459566073 (EA) 40198459566074 (CS), Lot Number 25JMF503; Medline Kit SKU DYNJ0376635Q: UDI/DI 10198459422591 (EA) 40198459422592 (CS), Lot Number 25KMH121; Medline Kit SKU DYNJ0376635Q: UDI/DI 10198459422591 (EA) 40198459422592 (CS), Lot Number 25KMF780; Medline Kit SKU DYNJ0460768T: UDI/DI 10198459466809 (EA) 40198459466800 (CS), Lot Number 25KMH789; Medline Kit SKU DYNJ0509186J: UDI/DI 10198459117626 (EA) 40198459117627 (CS), Lot Number 25KME895; Medline Kit SKU DYNJ0578815W: UDI/DI 10198459466892 (EA) 40198459466893 (CS), Lot Number 25JMJ695; Medline Kit SKU DYNJ0610290T: UDI/DI 10198459021558 (EA) 40198459021559 (CS), Lot Number 25KMC723; Medline Kit SKU DYNJ0753436P: UDI/DI 10198459466854 (EA) 40198459466855 (CS), Lot Number 25KMJ740; Medline Kit SKU…
    FDA recall number
    Z-1193-2026
    Quantity
    857 units

    The text above is the beginning of the record; 6,635 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Medline medical procedure convenience kits labeled as: HEAD SHAVE-LATEX SAFE KIT, Kit SKU P903933B

    Lot or code
    Medline Kit SKU P903933B: UDI/DI 10889942362080 (EA) 40889942362081 (CS), Lot Number 25KMA871
    FDA recall number
    Z-1194-2026
    Quantity
    60 units
  • Medline medical procedure convenience kits labeled as: 1) INPATIENT CTRL LN DRG CHG, Kit SKU DT22880; 2) ADULT CVC DRSG CHG W/ADHESIVE, Kit SKU DYNDC3283A.

    Lot or code
    Medline Kit SKU DT22880: UDI/DI 10653160995470 (EA) 40653160995471 (CS), Lot Number 25KMJ441; Medline Kit SKU DYNDC3283A: UDI/DI 10198459313240 (EA) 40198459313241 (CS), Lot Number 25KMH515.
    FDA recall number
    Z-1195-2026
    Quantity
    1,928 total
  • Medline medical procedure convenience kits labeled as: 1) DEPAUL BASIC NEURO-LF, Kit SKU DYNJ20028L; 2) NEURO MINOR PACK, Kit SKU DYNJ46546M; 3) NEURO SHUNT PACK, Kit SKU DYNJ54396K; 4) NEURO PACK, Kit SKU DYNJ59397Q; 5) NEURO BASIC, Kit SKU DYNJ904052L; 6) PK NEURO SHUNT BUMC, Kit SKU DYNJT6428.

    Lot or code
    Medline Kit SKU DYNJ20028L: UDI/DI 10198459228995 (EA) 40198459228996 (CS), Lot Number 25LMA824; Medline Kit SKU DYNJ46546M: UDI/DI 10198459526404 (EA) 40198459526405 (CS), Lot Number 25KMH940; Medline Kit SKU DYNJ54396K: UDI/DI 10198459443138 (EA) 40198459443139 (CS), Lot Number 25KMD804; Medline Kit SKU DYNJ59397Q: UDI/DI 10198459495427 (EA) 40198459495428 (CS), Lot Number 25KMC287; Medline Kit SKU DYNJ904052L: UDI/DI 10198459231339 (EA) 40198459231330 (CS), Lot Number 25JDB033; Medline Kit SKU DYNJT6428: UDI/DI 10198459553660 (EA) 40198459553661 (CS), Lot Number 25KMD463.
    FDA recall number
    Z-1196-2026
    Quantity
    236 units
  • Medline medical procedure convenience kits labeled as: 1) RR-GYN LAPAROSCOPY PACK-LF, Kit SKU DYNJ0395022O; 2) PELVISCOPY PACK-LF, Kit SKU DYNJ0415776Q; 3) DIVA PACK, Kit SKU DYNJ04811O; 4) DA VINCI PROSTATE/COLPOPEXY, Kit SKU DYNJ44864S; 5) MH ROBOTICS PACK, Kit SKU DYNJ50880L; 6) DELNOR LAPAROSCOPIC GYN PACK, Kit SKU DYNJ82244B; 7) ROBOT PACK, Kit SKU DYNJ82761B; 8) GENERAL ROBOTICS PACK, Kit SKU DYNJT4964.

    Lot or code
    Medline Kit SKU DYNJ0395022O: UDI/DI 10195327634827 (EA) 40195327634828 (CS), Lot Number 25KME682; Medline Kit SKU DYNJ0415776Q: UDI/DI 10198459566806 (EA) 40198459566807 (CS), Lot Number 25KMH125; Medline Kit SKU DYNJ04811O: UDI/DI 10198459440953 (EA) 40198459440954 (CS), Lot Number 25KMD772; Medline Kit SKU DYNJ44864S: UDI/DI 10195327127626 (EA) 40195327127627 (CS), Lot Number 25LMA115; Medline Kit SKU DYNJ50880L: UDI/DI 10198459388002 (EA) 40198459388003 (CS), Lot Number 25LMA516; Medline Kit SKU DYNJ82244B: UDI/DI 10198459344053 (EA) 40198459344054 (CS), Lot Number 25KMF328; Medline Kit SKU DYNJ82761B: UDI/DI 10198459039232 (EA) 40198459039233 (CS), Lot Number 25LMA391; Medline Kit SKU DYNJT4964: UDI/DI 10198459460098 (EA) 40198459460099 (CS), Lot Number 25KMJ341.
    FDA recall number
    Z-1197-2026
    Quantity
    1,928 total
  • Medline medical procedure convenience kits labeled as: CIRCUMCISION PACK, Kit SKU DYNJ16826O

    Lot or code
    Medline Kit SKU DYNJ16826O: UDI/DI 10195327412067 (EA) 40195327412068 (CS), Lot Number 25LMA655; Medline Kit SKU DYNJ16826O: UDI/DI 10195327412067 (EA) 40195327412068 (CS), Lot Number 25KMC911.
    FDA recall number
    Z-1198-2026
    Quantity
    1,928 total
  • Medline medical procedure convenience kits labeled as: 1) C-SECTION PACK, Kit SKU DYNJ35494C; 2) SMJ C-SECTION PACK-LF, Kit SKU DYNJ38582O; 3) RR-L&D C-SECTION PACK-LF, Kit SKU DYNJ82084F.

    Lot or code
    Medline Kit SKU DYNJ35494C: UDI/DI 10198459117879 (EA) 40198459117870 (CS), Lot Number 25KMD420; Medline Kit SKU DYNJ38582O: UDI/DI 10198459603709 (EA) 40198459603700 (CS), Lot Number 25KMI679; Medline Kit SKU DYNJ82084F: UDI/DI 10198459089206 (EA) 40198459089207 (CS), Lot Number 25KMD145.
    FDA recall number
    Z-1199-2026
    Quantity
    96 units
  • Medline medical procedure convenience kits labeled as: 1) WATKINS SPINE PACK-LF, Kit SKU DYNJ0275635M; 2) SM-SPINE PACK-LF, Kit SKU DYNJ0395037X; 3) TOTAL HIP PACK, Kit SKU DYNJ04809N; 4) MAIN/LAMINECTOMY PACK-LF, Kit SKU DYNJ0565499AN; 5) SCC/SPOC HAND & FOOT PACK, Kit SKU DYNJ27164Q; 6) ORTHO SPINE, Kit SKU DYNJ40221G; 7) URO GYN PACK, Kit SKU DYNJ45311J; 8) LAMINECTOMY PACK, Kit SKU DYNJ46330C; 9) DJ ORTHO SPINE, Kit SKU DYNJ48980M; 10) CERVICAL PACK, Kit SKU DYNJ52995F; 11) NORTH BACK FUSION PACK-LF, Kit SKU DYNJ56483L; 12) KNEE ARTHROSCOPY CHRISTUS, Kit SKU DYNJ61200B; 13) KNEE ARTHROSCOPY PACK-LF, Kit SKU DYNJ66150; 14) HIP PACK, Kit SKU DYNJ66630; 15) OHNS FREE FLAP A PACK, Kit SKU DYNJ69065F; 16) LUMBAR PACK, Kit SKU DYNJ69113C; 17) LAMINECTOMY PACK, Kit SKU DYNJ69401D; 18) LAMINECTOMY PACK, Kit SKU DYNJ81013D; 19) HIP/SHOULDER ARTHROSCOPY PACK, Kit SKU DYNJ84275A; 20) FLASC ACL PACK, Kit SKU DYNJ88971; 21) SPINE PACK MERCY GALENA, Kit SKU DYNJ89943; 22) SPINAL FUSION MT CARMEL, Kit SKU DYNJ900173J; 23) HAND, Kit SKU DYNJ904945G; 24) LAMINECTOMY, Kit SKU DYNJ909505D; 25) SHOULDER SPLIT, Kit SKU DYNJ910543F; 26) SHOULDER SPLIT, Kit SKU DYNJ910543G; 27) CRANIOTOMY PACK, Kit SKU DYNJT4724; 28) SPINE PACK, Kit SKU DYNJT5184.

    Lot or code
    Medline Kit SKU DYNJ0275635M: UDI/DI 10198459464799 (EA) 40198459464790 (CS), Lot Number 25KMH120; Medline Kit SKU DYNJ0395037X: UDI/DI 10198459600555 (EA) 40198459600556 (CS), Lot Number 25LMB119; Medline Kit SKU DYNJ04809N: UDI/DI 10198459440908 (EA) 40198459440909 (CS), Lot Number 25KMF969; Medline Kit SKU DYNJ0565499AN: UDI/DI 10195327318352 (EA) 40195327318353 (CS), Lot Number 25JMF339; Medline Kit SKU DYNJ0565499AN: UDI/DI 10195327318352 (EA) 40195327318353 (CS), Lot Number 25LMB019; Medline Kit SKU DYNJ27164Q: UDI/DI 10198459138102 (EA) 40198459138103 (CS), Lot Number 25KMJ742; Medline Kit SKU DYNJ40221G: UDI/DI 10198459448515 (EA) 40198459448516 (CS), Lot Number 25KME669; Medline Kit SKU DYNJ45311J: UDI/DI 10198459255670 (EA) 40198459255671 (CS), Lot Number 25KMI080; Medline Kit SKU DYNJ46330C: UDI/DI 10889942490066 (EA) 40889942490067 (CS), Lot Number 25KMD795; Medline Kit SKU DYNJ48980M: UDI/DI 10195327533694 (EA) 40195327533695 (CS), Lot Number 25KMC459; Medline Kit SKU DYNJ52995F: UDI/DI 10195327639778 (EA) 40195327639779 (CS), Lot Number 25KME064; Medline Kit SKU DYNJ56483L: UDI/DI 10195327518141 (EA) 40195327518142 (CS), Lot Number 25LMB049; Medline Kit SKU DYNJ61200B: UDI/DI 10195327604264 (EA) 40195327604265 (CS), Lot Number 25KBE867; Medline Kit SKU DYNJ66150: UDI/DI 10193489317558 (EA) 40193489317559 (CS), Lot Number 25KMC057; Medline Kit SKU DYNJ66630: UDI/DI 10193489368284 (EA) 40193489368285 (CS), Lot Number 25KMG435; Medline Kit SKU DYNJ69065F: UDI/DI…
    FDA recall number
    Z-1200-2026
    Quantity
    117 units

    The text above is the beginning of the record; 1,324 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): US Nationwide distribution.

Source

Issued by
U.S. Food and Drug Administration
Firm
Medline Industries, LP (other notices of this firm)
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Medline Industries, LP, Month: February 2026, Week 2026-W06. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 98226 on February 4, 2026.

Other notices of Medline Industries, LP