What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Software errors that can result in incorrect surgical instrument positioning during spinal surgery.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
5 more codes are in the record below.
Mazor X robotic guidance system REF: TPL0059
- Lot or code
- Software versions: 5.0.1, 5.1.2, 5.1.3/ UDI: 07290109180465, 07290109184524, 07290109181158, 00763000635169, 07290109183213, 07290109184517, 00763000419134, 00763000431761, 07290109184098, 07290109184838, 07290115751376, 07290115751895
- FDA recall number
- Z-1122-2026
- Quantity
- 549 systems
Distribution (FDA): Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, China, Croatia, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Malaysia, Mexico, Netherlands, Poland, Portugal, Puerto Rico, Romania, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, United States, Vietnam.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Mazor Robotics Ltd
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.