What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
7 more codes are in the record below.
MEDLINE beds labeled as: 1) Semi Electric Basic Homecare Bed, REF MDR107002E; 2) Semi Electric Basic Homecare Bed, 4-pack,REF MDR107002E-4
- Lot or code
- 1) UDI/DI 40080196320999, All lots; 2) UDI/DI 40080196320999, All Lots
- FDA recall number
- Z-1011-2026
- Quantity
- 373392 units
MEDLINE beds labeled as: Semi Electric Lightweight Homecare Bed, REF MDR107002L
- Lot or code
- UDI/DI 40080196294320, All lots
- FDA recall number
- Z-1012-2026
- Quantity
- 13245 units
MEDLINE beds labeled as: 1) Full Electric Basic Homecare Bed, REF MDR107003E; 2) Full Electric Basic Homecare Bed, 4-pack,REF MDR107003E-4.
- Lot or code
- 1) REF MDR107003E, UDI/DI 40080196321002, All lots; 2) REF MDR107003E-4, UDI/DI 40080196321002, All lots
- FDA recall number
- Z-1013-2026
- Quantity
- 138412 units
MEDLINE beds labeled as: Full Electric Low Basic Homecare Bed, REF MDR107003ELO
- Lot or code
- UDI/DI 10080196546992, All lots
- FDA recall number
- Z-1014-2026
- Quantity
- 22565 units
MEDLINE beds labeled as: Full Electric Lightweight Homecare Bed, REF MDR107003L
- Lot or code
- UDI/DI 40080196294344, All lots
- FDA recall number
- Z-1015-2026
- Quantity
- 35694 units
MEDLINE beds labeled as: Full Electric Low Lightweight Homecare Bed, REF MDR107003LO
- Lot or code
- UDI/DI 40080196325901, All lots
- FDA recall number
- Z-1016-2026
- Quantity
- 15581 units
Distribution (FDA): US, CA, PA, GUAM, VI
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Medline Industries, LP (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.