Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 98122: Securitas Healthcare LLC: Arial 900 MHz Call Station and 1 more

FDA recall 98122 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

392 more codes are in the record below.

  • Arial 900 MHz Call Station, Model Number CSK200-900, with Firmware version: 1.6.3

    Lot or code
    All units of the Affected Device Models sold between January 1st, 2023 and June 26th, 2024. UDI-DI: 00840948102653; Serial Numbers: 000CCC0E68FE 000CCC0E6901 000CCC0E6907 000CCC0E68FF 000CCC0E6638 000CCC0E6905 000CCC0E68FD 000CCC0E6904 000CCC0E6908 000CCC0E69AD 000CCC0E6902 000CCC0E69AE 000CCC0E69A9 000CCC0E69A8 000CCC0E6639 000CCC0E6903 000CCC0E69AA 000CCC0E6900 000CCC0E69A6 000CCC0E6B16 000CCC0E6B0D 000CCC0E6B12 000CCC0E6B20 000CCC0E6B10 000CCC0E6B1C 000CCC0E6B21 000CCC0E6B14 000CCC0E6B0C 000CCC0E6B1A 000CCC0E6B11 000CCC0E6B1F 000CCC0E6B22 000CCC0E6B15 000CCC0E6B0E 000CCC0E6B0F 000CCC0E69A5 000CCC0E69AC 000CCC0E69A3 000CCC0E69AB 000CCC0E69A4 000CCC0E6B17 000CCC0E6B19 000CCC0E6B0B 000CCC0E6B1E 000CCC0E6B13 000CCC0E6B18 000CCC0E6B1D 000CCC0E6B1B 000CCC0E6ACA 000CCC0E6ACB 000CCC0E6ABF 000CCC0E6AC4 000CCC0E6AAF 000CCC0E6ABB 000CCC0E6AB8 000CCC0E6ABD 000CCC0E6AB7 000CCC0E6ABE 000CCC0E6AC2 000CCC0E6ABA 000CCC0E6A5D 000CCC0E6A74 000CCC0E6A61 000CCC0E6A57 000CCC0E6A58 000CCC0E6A5C 000CCC0E6A62 000CCC0E6A79 000CCC0E6A6F 000CCC0E6A5B 000CCC0E6A5E 000CCC0E6A59 000CCC0E6A78 000CCC0E6B2C 000CCC0E6B26 000CCC0E6B24 000CCC0E6B35 000CCC0E6B3A 000CCC0E6B28 000CCC0E6B2E 000CCC0E6B29 000CCC0E6B31 000CCC0E6B23 000CCC0E6B2A 000CCC0E6B34 000CCC0E6B38 000CCC0E6B36 000CCC0E6B27 000CCC0E6B33 000CCC0E6B39 000CCC0E6B25 000CCC0E6B2D 000CCC0E6B2B 000CCC0E6B32 000CCC0E6B2F 000CCC0E6B37 000CCC0E6B30 000CCC0E6A6C 000CCC0E6A6B 000CCC0E6948 000CCC0E6A67 000CCC0E6637 000CCC0E6946 000CCC0E6945 000CCC0E6947…
    FDA recall number
    Z-1031-2026
    Quantity
    5627 units

    The text above is the beginning of the record; 71,790 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Arial 900 MHz Call Station, Model Number CSK200-900MR, with Firmware version: 1.6.3

    Lot or code
    All units of the Affected Device Models sold between January 1st, 2023 and June 26th, 2024. UDI-DI: 00840948102660; Serial Numbers:000CCC0E6CC9 000CCC0E6CC6 000CCC0E6CCC 000CCC0E6CCE 000CCC0E6CC5 000CCC0E6CD0 000CCC0E6D1F 000CCC0E6D1B 000CCC0E6D24 000CCC0E6D1C 000CCC0E6D22 000CCC0E6D01 000CCC0E6D04 000CCC0E6D19 000CCC0E6D1A 000CCC0E6D1E 000CCC0E6CC8 000CCC0E6D23 000CCC0E6CCD 000CCC0E6D1D 000CCC0E6CCB 000CCC0E6CCA 000CCC0E6CC7 000CCC0E6D20 000CCC0E6CCF 000CCC0E627E 000CCC0E6286 000CCC0E628C 000CCC0E6282 000CCC0E627C 000CCC0E6279 000CCC0E6289 000CCC0E627A 000CCC0E628E 000CCC0E627D 000CCC0E6288 000CCC0E6280 000CCC0E6284 000CCC0E6DDB 000CCC0E6D2E 000CCC0E6D28 000CCC0E6DE0 000CCC0E6D27 000CCC0E6D2C 000CCC0E6D2F 000CCC0E6DE3 000CCC0E6D2D 000CCC0E6D30 000CCC0E6DE1 000CCC0E6DD9 000CCC0E6DDF 000CCC0E6D25 000CCC0E6DDE 000CCC0E6DE4 000CCC0E6D2A 000CCC0E6D2B 000CCC0E6D29 000CCC0E6DDC 000CCC0E6DDA 000CCC0E6D26 000CCC0E6D36 000CCC0E6D32 000CCC0E6D06 000CCC0E6D33 000CCC0E6D3C 000CCC0E6D38 000CCC0E6D3B 000CCC0E6D07 000CCC0E6D05 000CCC0E6D0B 000CCC0E6D0A 000CCC0E6D0C 000CCC0E6D03 000CCC0E6D37 000CCC0E6D31 000CCC0E6D39 000CCC0E6D35 000CCC0E6D3A 000CCC0E6D34 000CCC0E6D08 000CCC0E6341 000CCC0E6D02 000CCC0E6D09 000CCC0E5B9F 000CCC0E5BD5 000CCC0E5BE0 000CCC0E6719 000CCC0E676D 000CCC0E676E 000CCC0E6DE2 000CCC0E6DDD 000CCC0E6CBC 000CCC0E6CC0 000CCC0E6CBF 000CCC0E6CBB 000CCC0E6CB9 000CCC0E6CBA 000CCC0E6D6A 000CCC0E6CC2 000CCC0E6CBE 000CCC0E6D61 000CCC0E6D67 000CCC0E6D68 000CCC0E6D62 000CCC0E6CC1…
    FDA recall number
    Z-1032-2026
    Quantity
    1898 units

    The text above is the beginning of the record; 23,309 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide distribution -US Nationwide and the country of Canada.

Source

Issued by
U.S. Food and Drug Administration
Firm
Securitas Healthcare LLC
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Securitas Healthcare LLC, Month: January 2026, Week 2026-W03. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.