It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800
- FDA recall number
- Z-0923-2026
- Quantity
- 35
Distribution (FDA): U.S. Nationwide distribution.