What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
BVI Leos Laser and Endoscopy System. Model Number: OME6000U.
- Lot or code
- Model Number: OME6000U. UDI: (01)00886158037576(11)250513(21). Serial Numbers: 77UBV500003, 77UBV500004, 77UBV500005, 77UBV500006, 77UBV500010.
- FDA recall number
- Z-1008-2026
- Quantity
- 5 units
Distribution (FDA): US Nationwide distribution in the states of FL, MI, OK, TX, WI.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Beaver-Visitec International, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.