What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
The NorthStar OCT Navigation Surgical Technique Guide (STG) incorrectly identified both Medtronic s Solera and Medtronic s Infinity system toolcards as compatible with the NorthStar OCT Navigation Instruments. However, the NorthStar OCT Navigation Instruments are compatible for use with Medtronic s Solera toolcards only.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
112 more codes are in the record below.
Orthofix NorthStar OCT Navigation Surgical Technique Guide, and instruments: Navigation Driver, REF: PC2-400012; Navigation Straight Bone Probe, REF: PC2-400013; Navigation, 3.5mm Drill, REF: PC2-400135; Navigation, 4.0mm Drill, REF: PC2-400140; Navigation, 4.5mm Drill, PC2-400145; Navigation, 5.5mm Drill, REF: PC2-400155; Navigation, 3.5mm Tap, REF: PC2-400235; Navigation, 4.0mm Tap, REF: PC2-400240; Navigation, 4.5mm Tap, REF: PC2-400245; Navigation, 4.5mm Dual Lead Tap, REF: PC2-400345; Navigation, 5.5mm Dual Lead Tap, REF: PC2-400355
- Lot or code
- Surgical Technique Guide revisions: D0003926A, D0003926B, D0003926C, D0003926D. REF/UDI-DI/Lot: PC2-400012/10889981260033/MK81527A, MK81527ARR, MK81527ARRR, MK81574AR, MK81574AR1, MK93487A, MK93488ARR, MK93489A, MK93490A, MK93490ARR, MK93491AR, MK93491AR1, MK943488A, MK95882A, MK95882ARR, ML52083B, ML53136B, ML53188B, ML69777B, ML69778B, ML71257B, MM1001498D, MM1001499D, MM1001748D, MM1001749D, MM1001813D, MM1001902D, MM1001903D, MM1002025D, MM1002316D, MM1002328D, MM1002344D; PC2-400013/10889981260040/AL657979C, AL658549C, AL658549CRRR, AL658641C, AL661470C, AL671917D, AL683456F, AL702785F; PC2-400135/10889981260057/BK104899A, BK104899AR, BK104899ARR, BK104899ARRR, BK104899ARRR4, BK106822A, CR49020D, CR49023D, CR49060D, CR51550E, CR51972D, CR56472E; PC2-400140/10889981260064/BK104900A, BK104900AR, BK104900ARRR, BK106825A, BK107806A, BK107806ARRR, BK108366A, BK108366ARRR, CR49822D, CR50158D, CR51973D, CR57547E, CR59411E; PC2-400145/10889981260071/BK104901A, BK104901ARRR, BK106827A, CR49021C, CR49061C, CR51974C; PC2-400155/10889981260088/BK104902A, BK104902ARRR, BK106829A, CR49062C, CR51551C, CR51975C, CR57550E, CR59413E; PC2-400235/10889981260095/BK104903A, BK104903ARRR, BK106831A, BK106831ARRR, BK131159C, CR49620C, CR52371D, CR54729D, CR58013D, CR59414D; PC2-400240/10889981260101/BK104904A, BK104904ARRR, BK106833A, BK131160C, CR49621C, CR52372D, CR59415D; PC2-400245/10889981275495/no lots released; PC2-400345/10889981260118/BK104905A, BK104905ARRR, BK106834A, BK106834ARRR,…
- FDA recall number
- Z-0937-2026
- Quantity
- 310
The text above is the beginning of the record; 108 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Distribution (FDA): US: CT, TX, NV, CO, CA, WA, NJ, OR, IN, MI, FL, PA, IL, VA, AL, RI, MO, AZ, MS, SC, UT, NY, OK, SD. OUS: Australia, Mexico, Chile
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- SEASPINE ORTHOPEDICS CORPORATION
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.