Why it was recalled
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing BD Luer Tip Caps. BD has issued a recall of their Luer Tip Caps after confirming through internal testing that certain BD Luer Tip Cap trays failed routine biocompatibility testing. Although the tip cap covers a small surface area of the syringe and does not come in direct contact with patients, there is the potential that the failure of a biocompatibility test may have varying consequences ranging from no harm or clinical impact to potential for sensitization, irritation, pyrogenicity, acute systemic toxicity or hemolysis depending on the nature and duration of contact with the syringe.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
MEDLINE PERFUSION SYRINGE, REF: DYNJ39223C
- Lot or code
- UDI/DI 10193489621532 (each) 40193489621533 (case), Lot Number 25FMJ081
- FDA recall number
- Z-0886-2026
- Quantity
- 15 units
Distribution (FDA): US Nationwide distribution in the state of CA.