Why it was recalled
Fibula nail system intended for use in fixation of fibular fractures and osteotomies, has out of specification thread depths on the fibula nail, so implant may not be able to mate properly with the associated mounting bolt. If nail (implant) cannot properly mate with instrumentation, the surgery potentially will need to be aborted.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Phantom Fibula Nail, 2.8mm x 130mm, Right, Ti, Fibula Nail, REF: P36-128-130R-S
- Lot or code
- UDI-DI: 00889795133670. Lot: 501182207125. Expiration: March 13th, 2030
- FDA recall number
- Z-0871-2026
- Quantity
- 20
Distribution (FDA): US Nationwide distribution in the states of NJ, WA, NY, MD, CA, FL, NM, TX, IL, OH, NV, TN, LA.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Paragon 28, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.