What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results. This error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes. The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results, this error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
9 more codes are in the record below.
FlexLab (FLX); Version: FLX-217-10;
- Lot or code
- UDI-DI: 07640172340004 Serial number: FLX.0226;
- FDA recall number
- Z-0595-2026
- Quantity
- 1 unit (OUS only)
FlexLab (FLX); Version: FLX-217-40;
- Lot or code
- UDI-DI: 07640172344002; Serial number: FXX.0004, FXX.0006, FXX.0009, FXX.0022, FXX.0023, FXX.0024, FXX.0027, FXX.0028, FXX.0034, FXX.0036, FXX.0041, FXX.0042, FXX.0043, FXX.0046;
- FDA recall number
- Z-0596-2026
- Quantity
- 14 units (1 US, 13 OUS)
Distribution (FDA): US Domestic: MI. OUS International: Australia, Finland, France, Germany, Italy, Portugal, Spain, and UK.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Inpeco S.A.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.