What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
219 more codes are in the record below.
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Product description/Product numbers: ADULT 120 EXPANDABLE W/MASK DYNJAA10591; ANES CIRC 120IN EXP-LF DYNJAA6555G; ANES CIRCUIT ADULT 120"-LF DYNJAA10817; ANESTHESIA 120" ADLT-LF DYNJAA10775; ANESTHESIA CIRC ADULT 120-LF DYNJAA10601; ANESTHESIA CIRCUIT - LF DYNJAA10261B; ANESTHESIA CIRCUIT 120EXP-LF DYNJAA10741A; ANESTHESIA CIRCUIT ADULT-LF DYNJAA0380A DYNJAA10641 DYNJAA10706B DYNJAA10851 DYNJAA10922B DYNJAA10971A DYNJAA11030A DYNJAA11087 DYNJAA11126; ANESTHESIA CIRCUIT-LF DYNJAA0309A DYNJAA0368A DYNJAA10368A DYNJAA10392 DYNJAA10430A DYNJAA10547A DYNJAA10639A DYNJAA10656A DYNJAA10685A DYNJAA10685B; CIRC 120" EXP 1 BV HMEF 3L GSL DYNJAA10920; EXPANDABLE CIRCUIT-LF DYNJAA11095; NM ANESTHESIA TURNOVER KIT DYNJAA0433; PROV HEALTH HEPA CIRC #2-LF DYNJAA0360; SPTMPX01IA CIRCUIT ADULT-LF DYNJAA10973A;
- Lot or code
- DYNJAA0309A UDI-DI 40193489400268 (case), 10193489400267 (ea) Lots 25EBU353 25GBI473; DYNJAA0360 UDI DI 40193489419000 (case), 10193489419009 (ea) Lots 25EBU304 25GBQ732; DYNJAA0368A UDI-DI 40195327092550 (case), 10195327092559 (ea) Lots 25GBH582; DYNJAA0380A UDI-DI 40195327478286 (case), 10195327478285 (ea) Lots 25FBB950 25FBT875 25GBX623; DYNJAA0433 UDI-DI 40198459320706 (case), 10198459320705 (ea) Lots 25EBH621 25EBQ391; DYNJAA10261B UDI- DI 40193489430784 (case), 10193489430783 (ea) Lots 25FBJ760; DYNJAA10368A UDI-DI 40198459320720 (case), 10198459320729 (ea) Lots 25GBY374 25IBK072; DYNJAA10392 UDI-DI 40888277962201 (case), 10888277962200 (ea) Lots 25FBN032; DYNJAA10430A UDI-DI 40193489832052 (case), 10193489832051 (ea) Lots 25FBW335 25IBH048; DYNJAA10547A UDI-DI 40193489603188 (case), 10193489603187 (ea) 25FBT853; DYNJAA10591 UDI-DI 40889942632535 (case), 10889942632534 (ea) Lots 25FBD876 25FBD877 25GBJ254 25GBJ256; DYNJAA10601 UDI-DI 40889942655435 (case), 10889942655434 (ea) Lots 25FBN184 25HBM893; DYNJAA10639A UDI-DI 40888277761958 (case), 10888277761957 (ea) Lots 25FBN043; DYNJAA10641 UDI-DI 40889942841227 (case), 10889942841226 (ea) Lots 25EBK360; DYNJAA10656A UDI-DI 40193489455527 (case), 10193489455526 (ea) Lots 25FBN055; DYNJAA10685A UDI-DI 40195327279142 (case), 10195327279141 (ea) Lots 25EBJ456; DYNJAA10685B UDI-DI 40198459372293 (case), 10198459372292 (ea) Lots 25FBL121 25FBT898 25FBT899; DYNJAA10706B UDI-DI 40193489485982 (case), 10193489485981 (ea) Lots…
- FDA recall number
- Z-0442-2026
- Quantity
- 96,406 total units
The text above is the beginning of the record; 1,141 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; BREATHING KIT-LF DYNJ10550
- Lot or code
- DYNJ10550 UDI-DI 40889942469704 (case), 10889942469703 (ea) Lot 25EBQ445
- FDA recall number
- Z-0443-2026
- Quantity
- 96,406 total units
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; OPEN HEART SUPPLEMENT DYNJ909116D; VASCULAR DYNJ909124F
- Lot or code
- DYNJ909116D UDI-DI 40198459055356 (case), 10198459055355 (ea) Lots 25FBM604 25GBP604; DYNJ909124F UDI-DI 40198459130619 (case), 10198459130618 (ea) Lots 25FBB831 25GBI833
- FDA recall number
- Z-0444-2026
- Quantity
- 96,406 total units
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; ENT DYNJ909128I; LARYNGOSCOPY - BRONCHOSCOPY DYNJ909114D
- Lot or code
- ENT DYNJ909128I UDI-DI 40198459333355 (case), 10198459333354 (ea) Lots 25GBO685; LARYNGOSCOPY - BRONCHOSCOPY DYNJ909114D UDI-DI 40198459055301 (case), 10198459055300 (ea) Lots 25GBN602
- FDA recall number
- Z-0445-2026
- Quantity
- 96,406 total units
DYNJ909119G ROBOTIC GASTRIC LAP SLEEVE-LF DYNJ909127F CYSTO DYNJ909129D GENERAL LAPAROSCOPY-LF
- Lot or code
- CYSTO DYNJ909127F UDI-DI 40198459055424 (case), 10198459055423 (ea) Lots 25FBP819 25GBM605; GENERAL LAPAROSCOPY-LF DYNJ909129D UDI-DI 40198459055264 (case), 10198459055263 (ea) Lots 25FBV537; ROBOTIC GASTRIC LAP SLEEVE-LF DYNJ909119G UDI-DI 40198459130527 (case), 10198459130526 (ea) Lots 25FBP681
- FDA recall number
- Z-0446-2026
- Quantity
- 96,406 total units
Convenience Kit, BREAST HERNIA DYNJ909125G
- Lot or code
- DYNJ909125G UDI-DI 40198459130664 (case), 10198459130663 (ea) Lot 25FBU543 25IBD459
- FDA recall number
- Z-0447-2026
- Quantity
- 96,406 total units
Convenience Kits: DYNJ901048N ORTHO SPINE ST CLAIR DYNJ909123F UROLOGY MINOR DYNJ909132F SET UP DYNJ909308D HIP PINNING
- Lot or code
- HIP PINNING DYNJ909308D UDI-DI 40198459055288 (case), 10198459055287 (ea) Lots 25FBB311 25GBC282 25GBW486; ORTHO SPINE ST CLAIR DYNJ901048N UDI-DI 40198459055554 (case), 10198459055553 (ea) Lots 25GBQ070; SET UP DYNJ909132F UDI-DI 40198459130572 (case), 10198459130571 (ea) Lots 25GBI845 25IBF708; UROLOGY MINOR DYNJ909123F UDI-DI 40198459130602 (case), 10198459130601 (ea) Lots 25GBT826 25IBF707
- FDA recall number
- Z-0448-2026
- Quantity
- 96,406 total units
Anesthesia circuit kit; CIRCUIT,ANES,EXTENDFLEX,120,LF DYNJAA10230
- Lot or code
- CIRCUIT,ANES,EXTENDFLEX,120,LF DYNJAA10230 UDI-DI 40884389688589 (case), 10884389688588 (ea) Lots 25EBV570 25FBT843
- FDA recall number
- Z-0449-2026
- Quantity
- 96,406 total units
Convenience kits; DYNJ901046M GYN ROBOTIC DYNJ909120D ROBOTIC PROSTATE
- Lot or code
- GYN ROBOTIC DYNJ901046M UDI-DI 40198459130381 (case), 10198459130380 (ea) Lots 25FBM810 25GBK307; ROBOTIC PROSTATE DYNJ909120D UDI-DI 40198459130541 (case), 10198459130540 (ea) Lots 25GBT821 25IBF809
- FDA recall number
- Z-0450-2026
- Quantity
- 96,406 total units
Convenience kits; DYNJ901047R ORTHO SPINE FUMICH DYNJ902308K TOTAL JOINT DYNJ909111F DR SALMA'S SPINE DYNJ909117G PATTERSON TOTAL JOINT DYNJ909113F KYPHOPLASTY
- Lot or code
- DR SALMA'S SPINE DYNJ909111F UDI-DI 40198459130794 (case), 10198459130793 (ea) Lots 25FBE036 25GBV759; ORTHO SPINE FUMICH DYNJ901047R UDI-DI 40198459130466 (case), 10198459130465 (ea) Lots 25FBR230 25GBU693; PATTERSON TOTAL JOINT DYNJ909117G UDI-DI 40198459130480 (case), 10198459130489 (ea) Lots 25FBO307 25GBN625; TOTAL JOINT DYNJ902308K UDI-DI 40198459130596 (case), 10198459130595 (ea) Lots 25FBB321 25GBU688 KYPHOPLASTY DYNJ909113F UDI-DI 40198459130459 (case), 10198459130458 (ea) Lot 25IBM788
- FDA recall number
- Z-0451-2026
- Quantity
- 96,406 total units
Distribution (FDA): International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Medline Industries, LP (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.