What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT;
- Lot or code
- CFN: SIGSDL45CTVT; UDI-DIs: (1) 20884521741840, (2) 10884521741843; Lot numbers: N5F1971Y, N5F1723Y, N5D1858UY, N5B1134UY;
- FDA recall number
- Z-0485-2026
- Quantity
- 3263 units
Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); BOX LOBECTOMIA VATS 3D; CFN: BOX06170V4;
- Lot or code
- CFN: BOX06170V4; UDI-DI: 00763000988975; Lot numbers: 0231961695;
- FDA recall number
- Z-0486-2026
- Quantity
- 2 units (OUS only)
Distribution (FDA): Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Chile, Estonia, Finland, France, Germany, Greece, Hungary, Israel, Italy, Japan, Kuwait, Malaysia, Netherlands, Poland, Portugal, Romania, Saudi Arabia, Slovakia, Spain, United Kingdom.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Covidien, LP
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.