Why it was recalled

Software issue for hq analyzer results in system not visibly applying appropriate flagging to results when saturation is present which may result in incorrect results.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

  • Brand Name: Alinity hq Analyzer Product Name: Alinity hq Analyzer Model/Catalog Number: 09P68-01 Software Version: 5.0 to 5.10.1 for OUS, 5.7 and 5.10.1 in the US Product Description: Alinity hq Analyzer Component: No

    Lot or code
    SN UDI HQ00480 (01)380740138851(21)HQ00480 HQ01231 (01)380740138851(21)HQ01231 HQ01232 (01)380740138851(21)HQ01232 HQ01233 (01)380740138851(21)HQ01233 HQ01236 (01)380740138851(21)HQ01236 HQ01237 (01)380740138851(21)HQ01237 HQ01238 (01)380740138851(21)HQ01238 HQ01239 (01)380740138851(21)HQ01239 HQ01240 (01)380740138851(21)HQ01240 HQ01241 (01)380740138851(21)HQ01241 HQ01244 (01)380740138851(21)HQ01244 HQ01246 (01)380740138851(21)HQ01246 HQ01247 (01)380740138851(21)HQ01247 HQ01248 (01)380740138851(21)HQ01248 HQ01249 (01)380740138851(21)HQ01249 HQ01250 (01)380740138851(21)HQ01250 HQ01252 (01)380740138851(21)HQ01252 HQ01253 (01)380740138851(21)HQ01253 HQ01256 (01)380740138851(21)HQ01256 HQ01257 (01)380740138851(21)HQ01257 HQ01263 (01)380740138851(21)HQ01263 HQ01265 (01)380740138851(21)HQ01265 OUS: HQ00117 (01)380740138851(21)HQ00117 HQ00119 (01)380740138851(21)HQ00119 HQ00120 (01)380740138851(21)HQ00120 HQ00126 (01)380740138851(21)HQ00126 HQ00127 (01)380740138851(21)HQ00127 HQ00128 (01)380740138851(21)HQ00128 HQ00132 (01)380740138851(21)HQ00132 HQ00136 (01)380740138851(21)HQ00136 HQ00139 (01)380740138851(21)HQ00139 HQ00140 (01)380740138851(21)HQ00140 HQ00141 (01)380740138851(21)HQ00141 HQ00144 (01)380740138851(21)HQ00144 HQ00145 (01)380740138851(21)HQ00145 HQ00148 (01)380740138851(21)HQ00148 HQ00149 (01)380740138851(21)HQ00149 HQ00152 (01)380740138851(21)HQ00152 HQ00154 (01)380740138851(21)HQ00154 HQ00157 (01)380740138851(21)HQ00157 HQ00162 (01)380740138851(21)HQ00162 HQ00164…
    FDA recall number
    Z-0489-2026
    Quantity
    833

    The text above is the beginning of the record; 28,504 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide distribution: US (nationwide) to states of: AL, CA, CO, FL, IN, MN, NC, NY, OK and OUS (foreign) to countries of: ALBANIA, ARGENTINA, AUSTRALIA, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CAYMAN ISLANDS, CHILE, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, INDIA, INDONESIA, ITALY, JAMAICA, JAPAN, JORDAN, KUWAIT, LEBANON, LITHUANIA, MALAYSIA, MALDIVES, MEXICO, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PANAMA, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, SAUDI ARABIA, SERBIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWITZERLAND, THAILAND, TRINIDAD & TOBAGO, TURKEY, UAE,UNITED KINGDOM, URUGUAY, and VIETNAM.

Source

Issued by
U.S. Food and Drug Administration
Firm
Abbott Laboratories
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Abbott Laboratories, Month: November 2025, Week 2025-W47. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Software issue for hq analyzer results in system not visibly applying appropriate flagging to results when saturation is present which may result in incorrect results.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 97788 on November 19, 2025.

Other notices of Abbott Laboratories