What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Terminated. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
SuRgical Planner (SRP), software version SRP.7.12.0, imaging processing radiological system
- Lot or code
- UDI/DI 00863157000336
- FDA recall number
- Z-0319-2026
- Quantity
- 45 units
Surgical Navigation Advanced Platform (SNAP), software version SNP.3.12.0, imaging processing radiological system
- Lot or code
- UDI/DI 00863157000305
- FDA recall number
- Z-0320-2026
- Quantity
- 17 units
Surgical Navigation Advanced Platform Patient Engagement (PE SNAP), software version SNP.3.12.0, imaging processing radiological system
- Lot or code
- UDI/DI 00863157000374
- FDA recall number
- Z-0321-2026
- Quantity
- 50 units
SyncAR Surgical Navigation Advanced Platform Patient Engagement (SyncAR SNAP), software version SNP.1.12.0, imaging processing radiological system
- Lot or code
- UDI/DI 00860003884102
- FDA recall number
- Z-0322-2026
- Quantity
- 14 units
Endo Surgical Navigation Advanced Platform Patient Engagement (EndoSNAP), software version SNP.1.12.0, imaging processing radiological system
- Lot or code
- UDI/DI 00863157000329
- FDA recall number
- Z-0323-2026
- Quantity
- 19 units
SpineAR Surgical Navigation Advanced Platform Patient Engagement (SyncAR SNAP), software version SPR.1.2.0, orthopedic augmented reality
- Lot or code
- UDI/DI 00860003884126
- FDA recall number
- Z-0324-2026
- Quantity
- 3 units
Surgical Planner, BrainSTORM, software version BST.9.3.0, orthopedic augmented reality
- Lot or code
- UDI/DI 00860003884119
- FDA recall number
- Z-0325-2026
- Quantity
- 3 units
Distribution (FDA): Worldwide - US Nationwide distribution in the states of MI, FL, DC, MN, CA, TX, NJ, PA, AZ, NY, KY, IN, WV and the countries of Italy, Germany, Israel, UK.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Surgical Theater Inc
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.