What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migration, subsidence and/or neurological injury. Therefore, surgical technique and instructions for use will be updated to mitigate risk of loss of lordosis to occur.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
392 more codes are in the record below.
Catalyft PL & PL40 Expandable Interbody System labeled as follows with corresponding model numbers: 1. SPACER 6068073 CATALYFT PL SHORT 7MM. Model Number: 6068073. 2. SPACER 6068076 CATALYFT PL LONG 7MM, Model Number: 6068076. 3. SPACER 6068093 CATALYFT PL SHORT 9MM, Model Number: 6068093. 4. SPACER 6068096 CATALYFT PL LONG 9MM, Model Number: 6068096. 5. SPACER 6068113 CATALYFT PL SHORT 11MM, Model Number: 6068113. 6. SPACER 6068116 CATALYFT PL LONG 11MM, Model Number: 6068116. 7. SPACER 6069073 CATALYFT PL40 SHORT 7MM, Model Number: 6069073. 8. SPACER 6069076 CATALYFT PL40 LONG 7MM, Model Number: 6069076. 9. SPACER 6069093 CATALYFT PL40 SHORT 9MM, Model Number: 6069093. 10. SPACER 6069096 CATALYFT PL40 LONG 9MM, Model Number: 6069096. 11. SPACER 6069113 CATALYFT PL40 SHORT 11MM, Model Number: 6069113. 12. SPACER 6069116 CATALYFT PL40 LONG 11MM, Model Number: 6069116. 13. CATALYFT PL IMPLANT SET, Model Number: SPS03143. 14. CATALYFT PL 40 IMPLANT SET, Model Number: SPS03144. 15. KIT CATALYFT PL IMPLANTS, Model Number: SPS04012. 16. KIT CATALYFT PL 40 IMPLANTS, Model Number: SPS04013. 17. CATALYFT PL 40 IMP KIT, Model Number: SPSJEP40. 18. CATALYFT PL 40 HIGH IMP KIT, Model Number: SPSJEP4H. 19. CATALYFT PL 40 SHORT IMP KIT, Model Number: SPSJEP4S. 20. CATALYFT PL 40 SHORT HIGH IMP KIT, Model Number: SPSJEP4T. 21. CATALYFT PL STRAIGHT HIGH IMP KIT, Model Number: SPSJEPSH. 22. CATALYFT PL STRAIGHT IMP KIT, Model Number: SPSJEPSI. These systems utilize the following affected…
- Lot or code
- IFU: M333023W048E Rev. A Surgical TechniqueM333023W217, Rev. 2.0 1. Model Number: 6068073. GTIN (Lot Numbers): 00763000246693 (0835083W, 0836192W, 0837725W, 0837727W, 0839577W, 0839579W, 0839591W, 0839593W, 0839595W, 0841459W, 0842257W, 0842767W, 0842769W, 0845001W, 0845003W, 0846533W, 0849546W, 0849548W, 0849998W, 0854451W, 0856696W, 0856698W, 0858703W, 0868179W, 0868231W, 0869637W, 0869695W, 0872240W, 0873202W, 0874502W, 0874504W, 0877435W, 0878670W, 0879342W, 0879887W, 0881316W, 0885958W, 0885960W, 0886068W, 0886070W, 0887342W, 0888799W, 0889934W, 0889936W, 0891419W, 0891951W, 0893270W, 0893332W, 0894905W, 0896952W, 0896954W, 0897308W, 0900773W, 0900970W, 0901192W, 0902362W, 0903362W, 0904350W, 0906154W, 0907656W, 0908818W, 0908820W, 0910150W, 0910202W, 0914572W, 0914574W, 0914576W, 0914578W, 0914580W, 0917478W, 0917480W, 0917662W, 0918071W, 0918073W, 0918075W, 0918077W, 0918858W, 0918860W, 0918862W, 0918864W, 0918866W, 0918886W, 0918888W, 0922033W, 0924471W, 0925397W, 0926139W, 0926361W, 0927266W, 0931914W, 0931916W, 0933069W, 0933071W, 0933073W, 0933075W, 0934823W, 0934825W, 0934827W, 0934998W, 0935000W, 0936452W, 0936454W, 0936456W, 0936458W, 0937870W, 0937872W, 0939113W, 0939115W, 0939324W, 0939326W, 0941080W, 0941364W, 0941366W, 0941912W, 0941914W, 0942912W, 0943255W, 0943257W, 0943953W, 0943955W, 0943957W, 0945701W, 0945946W, 0945948W, 0945950W, 0950146W, 0950148W, 0950150W, 0951512W, 0951514W, 0951516W, 0952750W, 0952752W, 0954062W, 0954064W, 0954430W, 0955352W,…
- FDA recall number
- Z-0580-2026
- Quantity
- 275,260 units
The text above is the beginning of the record; 48,988 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Distribution (FDA): Worldwide distribution - US Nationwide and Puerto Rico. The countries of Australia, Austria, Belgium, Canada, Croatia, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Netherlands, New Caledonia, New Zealand, Norway, Portugal, Reunion, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Medtronic Sofamor Danek USA Inc
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.