Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 97714: Medtronic Sofamor Danek USA Inc: Catalyft PL & PL40 Expandable Interbody System labeled as follows with corresponding…

FDA recall 97714 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migration, subsidence and/or neurological injury. Therefore, surgical technique and instructions for use will be updated to mitigate risk of loss of lordosis to occur.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

392 more codes are in the record below.

  • Catalyft PL & PL40 Expandable Interbody System labeled as follows with corresponding model numbers: 1. SPACER 6068073 CATALYFT PL SHORT 7MM. Model Number: 6068073. 2. SPACER 6068076 CATALYFT PL LONG 7MM, Model Number: 6068076. 3. SPACER 6068093 CATALYFT PL SHORT 9MM, Model Number: 6068093. 4. SPACER 6068096 CATALYFT PL LONG 9MM, Model Number: 6068096. 5. SPACER 6068113 CATALYFT PL SHORT 11MM, Model Number: 6068113. 6. SPACER 6068116 CATALYFT PL LONG 11MM, Model Number: 6068116. 7. SPACER 6069073 CATALYFT PL40 SHORT 7MM, Model Number: 6069073. 8. SPACER 6069076 CATALYFT PL40 LONG 7MM, Model Number: 6069076. 9. SPACER 6069093 CATALYFT PL40 SHORT 9MM, Model Number: 6069093. 10. SPACER 6069096 CATALYFT PL40 LONG 9MM, Model Number: 6069096. 11. SPACER 6069113 CATALYFT PL40 SHORT 11MM, Model Number: 6069113. 12. SPACER 6069116 CATALYFT PL40 LONG 11MM, Model Number: 6069116. 13. CATALYFT PL IMPLANT SET, Model Number: SPS03143. 14. CATALYFT PL 40 IMPLANT SET, Model Number: SPS03144. 15. KIT CATALYFT PL IMPLANTS, Model Number: SPS04012. 16. KIT CATALYFT PL 40 IMPLANTS, Model Number: SPS04013. 17. CATALYFT PL 40 IMP KIT, Model Number: SPSJEP40. 18. CATALYFT PL 40 HIGH IMP KIT, Model Number: SPSJEP4H. 19. CATALYFT PL 40 SHORT IMP KIT, Model Number: SPSJEP4S. 20. CATALYFT PL 40 SHORT HIGH IMP KIT, Model Number: SPSJEP4T. 21. CATALYFT PL STRAIGHT HIGH IMP KIT, Model Number: SPSJEPSH. 22. CATALYFT PL STRAIGHT IMP KIT, Model Number: SPSJEPSI. These systems utilize the following affected…

    Lot or code
    IFU: M333023W048E Rev. A Surgical TechniqueM333023W217, Rev. 2.0 1. Model Number: 6068073. GTIN (Lot Numbers): 00763000246693 (0835083W, 0836192W, 0837725W, 0837727W, 0839577W, 0839579W, 0839591W, 0839593W, 0839595W, 0841459W, 0842257W, 0842767W, 0842769W, 0845001W, 0845003W, 0846533W, 0849546W, 0849548W, 0849998W, 0854451W, 0856696W, 0856698W, 0858703W, 0868179W, 0868231W, 0869637W, 0869695W, 0872240W, 0873202W, 0874502W, 0874504W, 0877435W, 0878670W, 0879342W, 0879887W, 0881316W, 0885958W, 0885960W, 0886068W, 0886070W, 0887342W, 0888799W, 0889934W, 0889936W, 0891419W, 0891951W, 0893270W, 0893332W, 0894905W, 0896952W, 0896954W, 0897308W, 0900773W, 0900970W, 0901192W, 0902362W, 0903362W, 0904350W, 0906154W, 0907656W, 0908818W, 0908820W, 0910150W, 0910202W, 0914572W, 0914574W, 0914576W, 0914578W, 0914580W, 0917478W, 0917480W, 0917662W, 0918071W, 0918073W, 0918075W, 0918077W, 0918858W, 0918860W, 0918862W, 0918864W, 0918866W, 0918886W, 0918888W, 0922033W, 0924471W, 0925397W, 0926139W, 0926361W, 0927266W, 0931914W, 0931916W, 0933069W, 0933071W, 0933073W, 0933075W, 0934823W, 0934825W, 0934827W, 0934998W, 0935000W, 0936452W, 0936454W, 0936456W, 0936458W, 0937870W, 0937872W, 0939113W, 0939115W, 0939324W, 0939326W, 0941080W, 0941364W, 0941366W, 0941912W, 0941914W, 0942912W, 0943255W, 0943257W, 0943953W, 0943955W, 0943957W, 0945701W, 0945946W, 0945948W, 0945950W, 0950146W, 0950148W, 0950150W, 0951512W, 0951514W, 0951516W, 0952750W, 0952752W, 0954062W, 0954064W, 0954430W, 0955352W,…
    FDA recall number
    Z-0580-2026
    Quantity
    275,260 units

    The text above is the beginning of the record; 48,988 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide distribution - US Nationwide and Puerto Rico. The countries of Australia, Austria, Belgium, Canada, Croatia, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Netherlands, New Caledonia, New Zealand, Norway, Portugal, Reunion, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.

Source

Issued by
U.S. Food and Drug Administration
Firm
Medtronic Sofamor Danek USA Inc
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Medtronic Sofamor Danek USA Inc, Month: December 2025, Week 2025-W49. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migration, subsidence and/or neurological injury. Therefore, surgical technique and instructions for use will be updated to mitigate risk of loss of lordosis to occur.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 97714 on December 3, 2025.

Other notices of Medtronic Sofamor Danek USA Inc