Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 97668: UIH Technologies LLC: Shanghai United Imaging Healthcare Co. and 8 more

FDA recall 97668 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

154 more codes are in the record below.

  • Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 530 MD, Rx Only

    Lot or code
    Serial Numbers: ; UDI: (01)06971576831012/ Serial Numbers: 300162 300163 300174 300176 304002 304003 304004 304006
    FDA recall number
    Z-0217-2026
    Quantity
    8 units
  • Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 550 MD, Rx Only

    Lot or code
    UDI:(01)06971576831036/ Serial Numbers: 353020 353021 353023 353045 353049 354001 354002 354003 354004 354005 354006 354007 354008 354009 354010 354011 354012 354013 354014 354015 354016 354017
    FDA recall number
    Z-0218-2026
    Quantity
    22 units
  • Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 760 MD, Rx Only

    Lot or code
    UDI: (01)06971576831043/Serial Number: 600229 600232 600271 600282 604001 604002 604003 604004 604005 604006 604008
    FDA recall number
    Z-0219-2026
    Quantity
    11 units
  • Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 780 MD, Rx Only

    Lot or code
    UDI: (01)06971576831050/Serial Numbers: 684001, 684002
    FDA recall number
    Z-0220-2026
    Quantity
    2 units
  • Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT ATLAS MD, Rx Only

    Lot or code
    UDI: (01)06971576831104/ Serial Numbers: 863001, 863002
    FDA recall number
    Z-0221-2026
    Quantity
    2 units
  • Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uEXPLORER MD, Rx Only

    Lot or code
    UDI: (01)06971576832040; Serial Number: F00002, F00015
    FDA recall number
    Z-0222-2026
    Quantity
    2 UNITS
  • Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI 550 MD, Rx Only

    Lot or code
    UDI: (01)06971576832026/ Serial Numbers: 200017 200023 200024 200036 200045 200047 230002 230003 230004 230005 230006 230007 230008 230009 230011 230010 230012 230014 230013 200059 230015 230016 230017 230019 230020 230021 230022 230023 230024 230025 232001 200071 230026 230027 200072 230028 200074 230030 230029 200075 200076 230031 230032 230033 230034 230035 200079 200078 230037 230036 230038 230039 230040 230041 230042 230043 230044 230045 230047 230048 232002 202002 232003 232004 232005 230057 232006 232007 202004 232008 232009 202005 232010 202006 232011 202007 232012 232013 232014 230067 230068 230069 232015 232016 230070 230071 230073 202008 230072 230074 232017 202009 232018 232019 202010
    FDA recall number
    Z-0223-2026
    Quantity
    95 units
  • Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI 780 MD, Rx Only

    Lot or code
    UDI: (01)06971576832033/ Serial Number: 300072, 300082, 300098
    FDA recall number
    Z-0224-2026
    Quantity
    3 units
  • Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI Panorama 35 MD, Rx Only

    Lot or code
    UDI: (01)06971576832170/ Serial numbers: 11MI410003, 11MI410004, 11MI410005, 20MI410001
    FDA recall number
    Z-0225-2026
    Quantity
    4 unit

Distribution (FDA): US: OUS:

Source

Issued by
U.S. Food and Drug Administration
Firm
UIH Technologies LLC
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Month: October 2025, Week 2025-W44. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 97668 on October 29, 2025.