Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 97646: Wipro GE Healthcare Private Ltd.: GE Healthcare Carescape Central Station and 1 more

FDA recall 97646 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

391 more codes are in the record below.

  • GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; Medical Display, cathode ray tube

    Lot or code
    GTIN 00195278343512, Serial Numbers: STU25210418SA, STU25210417SA, STU25210419SA, STU25250440SA, STU25250439SA, STU25260447SA, STU25210407SA, STU25210406SA, STU24200003SA, STU24200006SA, STU24200005SA, STU24200004SA, STU23240015SA, STU24240019SA, STU24240020SA, STU24240007SA, STU24240022SA, STU24240021SA, STU24330020SA, STU24240014SA, STU24240013SA, STU24330019SA, STU24330022SA, STU24330016SA, STU24240005SA, STU24240008SA, STU24240012SA, STU24330018SA, STU24330021SA, STU24330017SA, STU23160011SA, STU23240007SA, STU24310001SA, STU25210433SA, STU25210432SA, STU25200390SA, STU25200392SA, STU25200391SA, STU25200399SA, STU25200397SA, STU25200395SA, STU25200393SA, STU25200398SA, STU25200396SA, STU25250438SA, STU25270475SA, STU25270456SA, STU25270468SA, STU25270458SA, STU25270466SA, STU25270476SA, STU25270469SA, STU25270473SA, STU25270459SA, STU25270474SA, STU25270467SA, STU25210430SA, STU25210429SA, STU25180343SA, STU25190347SA, STU25190348SA, STU24200007SA, STU25190360SA, STU25240436SA, STU25190361SA, STU25190362SA, STU25190363SA, STU25240437SA, STU25190359SA, STU24510027SA, STU24510032SA, STU24510010SA, STU24510020SA, STU24510017SA, STU24510026SA, STU24510018SA, STU24510024SA, STU24510025SA, STU24510031SA, STU24510011SA, STU24510029SA, STU24510023SA, STU24510030SA, STU24510021SA, STU24510019SA, STU24510022SA, STU25190350SA, STU25200388SA, STU25200389SA, STU25210412SA, STU25190349SA, STU25190381SA, STU25190376SA, STU25190383SA, STU25190374SA, STU25190367SA, STU25190372SA,…
    FDA recall number
    Z-0213-2026
    Quantity
    173 units

    The text above is the beginning of the record; 1,140 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5, Part Number 5867474-03; Medical Display, cathode ray tube

    Lot or code
    GTIN 00195278343383, Serial Numbers: STV25225456SA, STV25225455SA, STV25266464SA, STV25235817SA, STV25235816SA, STV25286682SA, STV25225594SA, STV25235792SA, STV25235774SA, STV25225550SA, STV25235626SA, STV25235625SA, STV25184592SA, STV25235628SA, STV25235631SA, STV25184595SA, STV25235629SA, STV25235632SA, STV25235630SA, STV25276515SA, STV25225510SA, STV25276516SA, STV25225505SA, STV25225508SA, STV25225509SA, STV25225507SA, STV25225506SA, STV25113369SA, STV25204986SA, STV25205160SA, STV25205151SA, STV25205161SA, STV25062736SA, STV25256133SA, STV25225512SA, STV25296702SA, STV25296701SA, STV25296748SA, STV25256214SA, STV25225511SA, STV25256215SA, STV25256199SA, STV25256216SA, STV25256202SA, STV25225497SA, STV25225514SA, STV25225498SA, STV25256131SA, STV25256129SA, STV25256128SA, STV25296747SA, STV25256132SA, STV25256362SA, STV25225494SA, STV25225513SA, STV25296697SA, STV25256153SA, STV25296745SA, STV25256130SA, STV25256134SA, STV25256198SA, STV25296749SA, STV25296698SA, STV25256112SA, STV25256111SA, STV25225495SA, STV25256127SA, STV25256197SA, STV25296700SA, STV25296699SA, STV23440159SA, STV23440160SA, STV23440161SA, STV23440162SA, STV23440163SA, STV23440164SA, STV23440165SA, STV23440166SA, STV23440167SA, STV23440169SA, STV23440170SA, STV23440172SA, STV23440173SA, STV23440174SA, STV23440175SA, STV23440176SA, STV23440178SA, STV23440179SA, STV23440180SA, STV23440185SA, STV23440186SA, STV23440187SA, STV23440188SA, STV24250149SA, STV24250150SA, STV24250151SA, STV23440171SA,…
    FDA recall number
    Z-0214-2026
    Quantity
    2365 units

    The text above is the beginning of the record; 34,111 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide distribution.

Source

Issued by
U.S. Food and Drug Administration
Firm
Wipro GE Healthcare Private Ltd.
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Month: October 2025, Week 2025-W44. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 97646 on October 29, 2025.