What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
The subject product emits unsafe level of UVC radiation (i.e., the product reaches the exposure limit for Risk Group 3 (high risk) in about 2.2 minutes). The product has a radiation safety defect, in that these radiation emissions are unnecessary to the accomplishment of the product s primary purpose of disinfecting objects and surfaces, and create a risk of skin and/or eye injuries to user and nearby persons.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
The product is a handheld ultraviolet-C germicidal wand and contains a narrow-diameter, tubular lamps with a peak emission wavelength in the 254 nm range.
- Lot or code
- SKU BT-GERMWAND-LRG-WHT and Testing: September 8-10, 2020
- FDA recall number
- Z-2571-2025
Distribution (FDA): U.S.A
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- BigTree Sales Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.