Skip to content
FDAMedical devicesClass IOngoing

FDA recall 97615: Cook Medical Incorporated: Zenith Alpha 2 Thoracic Endovascular Graft

FDA recall 97615 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Affected devices may contain PTFE coating scrapings. Scrapings could be released during device deployment, potentially causing intravascular embolization.
Text from FDA, unedited.

FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

392 more codes are in the record below.

  • Zenith Alpha 2 Thoracic Endovascular Graft, Catalog Prefix ZTA2- Reference Part Numbers ZTA2-P-40-117-W ZTA2-P-40-167 ZTA2-P-40-167-W ZTA2-P-40-217-W ZTA2-P-42-121 ZTA2-P-42-121-W ZTA2-P-42-173 ZTA2-P-42-173-W ZTA2-P-42-225 ZTA2-P-42-225-W ZTA2-P-44-125 ZTA2-P-44-125-W ZTA2-P-44-179 ZTA2-P-44-179-W ZTA2-P-44-233 ZTA2-P-44-233-W ZTA2-P-46-125 ZTA2-P-46-125-W ZTA2-P-46-179 ZTA2-P-46-179-W ZTA2-P-46-233-W ZTA2-PT-40-36-167 ZTA2-PT-40-36-167-W ZTA2-PT-40-36-217 ZTA2-PT-40-36-217-W ZTA2-PT-42-38-173 ZTA2-PT-42-38-173-W ZTA2-PT-42-38-225 ZTA2-PT-42-38-225-W ZTA2-PT-44-40-179 ZTA2-PT-44-40-179-W ZTA2-PT-44-40-233-W ZTA2-PT-46-42-179 ZTA2-PT-46-42-179-W ZTA2-PT-46-42-233 ZTA2-PT-46-42-233-W ZTA2-P-46-233 ZTA2-PT-44-40-233 ZTA2-P-40-217 ZTA2-P-40-117 ZTA2-P-40-167-DEMO ZTA2-P-44-152-W ZTA2-P-40-142 ZTA2-P-44-152 ZTA2-P-40-142-W

    Lot or code
    ZTA2-P-40-117-W Box UDI-DI 10827002599151 Product UDI-DI 00827002599154 Lots E4626917 E4627437 E4627438 E4627439 E4627440 E4629948 E4629949 E4629950 E4629951 E4629952 E4629953 E4629954 E4629955 E4634911 E4634913 E4634915 E4634916 E4634917 E4646219 E4656519 E4656520 E4656521 E4659972 E4659973 E4659974 E4665450 E4679401 E4679406 E4683233 E4683236 E4683237 E4683238 E4684592 E4684593 E4701373 E4701375 E4701376 E4701377 E4701378 E4701379 E4701380 E4701402 E4704235; ZTA2-P-40-167 Box UDI-DI 10827002599205 Product UDI-DI 00827002599208 Lots E4715752 E4718032 E4718033 E4718034 E4718035; ZTA2-P-40-167-W Box UDI-DI 10827002599212 Product UDI-DI 00827002599215 Lots E4627442 E4627443 E4627444 E4627445 E4627446 E4627447 E4627448 E4627449 E4627450 E4627451 E4627452 E4631968 E4631969 E4631970 E4634918 E4634919 E4634920 E4635173 E4635174 E4646220 E4656525 E4656526 E4656527 E4656528 E4656529 E4659975 E4665451 E4665452 E4665453 E4670244 E4670245 E4672425 E4672426 E4672427 E4675854 E4677105 E4679407 E4679408 E4679409 E4679410 E4679411 E4679412 E4679413 E4679414 E4679415 E4683243 E4684597 E4684598 E4684601 E4690174 E4690176 E4690178 E4690180 E4701381 E4701382 E4701383 E4701384 E4701385 E4701386 E4704236; ZTA2-P-40-217-W Box UDI-DI 10827002599236 Product UDI-DI 00827002599239 Lots E4629331 E4629332 E4629335 E4629337 E4629338 E4629339 E4646221 E4662229 E4662230 E4662231 E4665454 E4665455 E4665456 E4670246 E4670247 E4672428 E4672431 E4672432 E4675858 E4683246 E4683247 E4695558 E4704238 E4704239;…
    FDA recall number
    Z-0043-2026
    Quantity
    695

    The text above is the beginning of the record; 7,591 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Domestic distribution to AL AR AZ CA CT DC DE FL GA IL IN LA MA MD MI MN MO NC NJ NY OH OR PA SC TN TX UT VA WA WI WV International distribution to Switzerland, Germany, Italy, Poland, and Sweden

Source

Issued by
U.S. Food and Drug Administration
Firm
Cook Medical Incorporated
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Month: October 2025, Week 2025-W43. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Affected devices may contain PTFE coating scrapings. Scrapings could be released during device deployment, potentially causing intravascular embolization.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 97615 on October 22, 2025.