FDAMedical devicesClass IIOngoing

FDA recall 97580: MICROVENTION INC.: Flow Re-Direction Endoluminal Devices: FRED 27

FDA recall 97580 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Flow diverter may have a manufacturing issue related to the tantalum length and/or tantalum attachment pattern, which may lead the device to be incompletely open or not properly apposed to the vessel wall, which may necessitate device recapture and withdrawal of the delivery system and microcatheter together. Insufficient apposition could lead to long-term health consequences.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

92 more codes are in the record below.

  • Flow Re-Direction Endoluminal Devices: FRED 27, REF: FRED5019, FRED5029, FRED4528, FRED4026, FRED5526, FRED4508, FRED5514, FRED3536, FRED3511, FRED5014, FRED3516, MV-F451827; FRED X 27, REF: XFRED4017-MVE, XFRED4528-MVE, MV-F501427X, XFRED4518-MVE, XFRED5526-MVE, XFRED4539-MVE, FREDX5522-PMA, FREDX3522-PMA, FREDX4525-PMA, FREDX5015-PMA, FREDX5532-PMA, FREDX4013-PMA, FREDX3513-PMA, FREDX4515-PMA, FREDX3517-PMA, MV-F352427X

    Lot or code
    REF/UDI-DI(Lot): FRED5019/00811425020388(0000235839), 00842429117163(0000544498); FRED5029/00811425020395(0000256343), 00842429117170(0000507209); FRED4528/00811425020425(0000265666); FRED4026/00811425021033(0000333125, 0000576973); FRED5526/00842429117194(0000361117); FRED4508/00842429117095(0000372088); FRED5514/00842429117187(0000429786, 0000478385); FRED3536/00842429117033(0000460268, 0000445555); FRED3511/00842429117002(0000452923); FRED5014/00842429117156(0000563591, 0000464401); FRED3516/00842429117019(0000513146, 0000531733, 0000536217); MV-F451827/04987892121906(0000565094); XFRED4017-MVE/00842429107508(0000478838); XFRED4528-MVE/00842429107560(0000518323, 0000663751); MV-F501427X/04987892151231(0000579031); XFRED4518-MVE/00842429107553(0000663074, 0000663750); XFRED5526-MVE/00842429107638(0000704203); XFRED4539-MVE/00842429107577(00007311620); FREDX5522-PMA/00842429114483(0000265641, 0000469609, 0000731155); FREDX3522-PMA/00842429114315(0000469598); FREDX4525-PMA/00842429114414(0000469603); FREDX5015-PMA/00842429114445(0000469606); FREDX5532-PMA/00842429114490(0000469610); FREDX4013-PMA/00842429114346(0000469613, 0000663758, 0000668189); FREDX3513-PMA/00842429114292(0000478834); FREDX4515-PMA/00842429114391(0000478840); FREDX3517-PMA/00842429114308(0000731304); MV-F352427X/04987892151101(0000760963)
    FDA recall number
    Z-0108-2026
    Quantity
    467

Distribution (FDA): Worldwide - US Nationwide distribution in the states of SD, CA, SC, TX, FL, CO, NE, NY and the countries of DE, ES, PL, GB, IT, IN, AU, KZ, JP, AT, SI, RS, GR, TR, AR, CH, SG, KR, SK, MX, DK, IR, SA, BY, PA, VN, ZA, BR, TW, CN, BE, FR, VE, MY.

Source

Issued by
U.S. Food and Drug Administration
Firm
MICROVENTION INC.
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: MICROVENTION INC., Month: October 2025, Week 2025-W43. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Flow diverter may have a manufacturing issue related to the tantalum length and/or tantalum attachment pattern, which may lead the device to be incompletely open or not properly apposed to the vessel wall, which may necessitate device recapture and withdrawal of the delivery system and microcatheter together. Insufficient apposition could lead to long-term health consequences.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 97580 on October 22, 2025.