What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Production assembly error resulting in the potential for reversed articulation, where moving the lever forward caused the probe tip to anteflex rather than retroflex, and vice versa.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
ClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile Ultrasound, ImaCor model ZHH.
- Lot or code
- Model CLT-010; UDI-DI 00861589000108; Lot number: 20261209.
- FDA recall number
- Z-0038-2026
- Quantity
- 45 units
Distribution (FDA): US Nationwide distribution in the states of South Carolina, New Jersey, Pennsylvania, Florida, and Georgia.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- ImaCor Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.