What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Hemostasis valve has a manufacturing defect and when used the valve cap may remain depressed, leading to potential leakage, which may result in procedure inconvenience and/or hemorrhage.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
118 more codes are in the record below.
PhD Hemostasis Valve, REF: IN4802/K, IN4802/T, IN4918/A, IN7802/C, IN8802/C, IN9802, IN9802/A, K05-02166B, K05T-02138A, K12T-10798A, MAP800/B, MAP800/T, MAP801/B, MAP802/B, MAP802/CNB, MAP802/T, MAP804/B, MAP852/B, SKY1802.
- Lot or code
- REF/UDI-DI/Lot(Expiration): IN4802/K/00884450638385/I3138883(20-FEB-2028), I3150031(27-FEB-2028), I3161209(10-MAR-2028), I3161210(08-APR-2028), I3161211(17-MAR-2028), I3172451(30-NOV-2026); IN4802/T/00884450860113/I3135188(30-NOV-2026, I3170171(10-MAR-2028; IN4918/A/00884450638521/I3146050(20-FEB-2028), I3200779(19-MAY-2028); IN7802/C/00884450498743/H3143753(30-NOV-2026), H3162387(14-MAR-2028), H3183068(31-MAR-2028), H3189446(29-FEB-2028), H3213617(31-MAY-2028); IN8802/C/00884450621417/H3141570(02-FEB-2028), H3143769(03-FEB-2028), H3143770(31-DEC-2026), H3143771(03-FEB-2028), H3143772(03-FEB-2028), H3149478(17-FEB-2028), H3149480(31-DEC-2026), H3149483(31-DEC-2026), H3155549(24-FEB-2028), H3162422(14-MAR-2028), H3162627(14-MAR-2028), H3168247(07-MAR-2028), H3183054(31-DEC-2026), H3200495(26-APR-2028), H3219087(10-JUN-2028), H3241812(31-MAY-2028); IN9802/00884450413463/H3141573(21-MAR-2028), H3183200(28-MAR-2028); IN9802/A/00884450021347/H3200469(26-APR-2028), H3210336(30-NOV-2026); K05-02166B/00884450499597/T3192808(24-FEB-2027), T3198985(11-AUG-2027), T3201276(27-FEB-2027), T3217490(02-SEP-2027); K05T-02138A/ 00884450095607 T3232335(15-AUG-2027); K12T-10798A/00884450544372/T3163727(21-MAY-2027), T3217266(20-FEB-2026); MAP800/B/ 00884450746714/I3124454(07-JAN-2028), I3174736(10-MAR-2028), I3186951(24-MAR-2028), I3191334(01-APR-2028), I3206933(21-APR-2028), I3209939(29-APR-2028), I3219784(12-MAY-2028), I3219792(12-MAY-2028), I3234557(19-MAY-2028), I3234558(26-MAY-2028);…
- FDA recall number
- Z-0533-2026
- Quantity
- 217,141
The text above is the beginning of the record; 1,411 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Distribution (FDA): US: PA, AZ, LA, CO, FL, NC, SC, OH, AK, IL, IA, MN, TX, WI, AR, NY, CA, KS, NJ, MI, VA, KY, AL, MT, NE, MA, MO, TN, CT, MD, WA, DC, ME, OR, NH, IN, GA, UT, OK, HI, RI, ID, NM, MS, WV, SD, DE, NV, ND. OUS: United Arab Emirates, Australia, Italy, Spain, Sweden, Norway, Germany, France, Belgium, Netherlands, Thailand, New Zealand, Mexico, Puerto Rico, Canada, Korea, Republic of, South Africa, Qatar, Iceland, United Kingdom, Viet Nam, Slovakia, Guam, Oman, Portugal, Madagascar, Saudi Arabia, Finland, Cyprus, Switzerland, Taiwan, China, Japan
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Merit Medical Systems, Inc. (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.