Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 97429: Merge Healthcare, Inc.: Merge Hemo

FDA recall 97429 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Merge Healthcare has been informed by Schiller AG (Baar, Switzerland) of an issue with the Schiller ARGUS PB-3000 Patient Data Module (PDM) where certain units trigger an error message on the Merge Hemo Client (Pressure Sensor Defective) during non-invasive blood pressure (NIBP) measurements, specifically when the initial inflation pressure is set in the high-pressure range (>240 mmHg). The ARGUS PB-3000 is a vital signs monitoring instrument intended to be integrated into the Merge Hemo System. No other components of the Merge Hemo System are affected.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

80 more codes are in the record below.

  • Merge Hemo, Model RCSV2, Model/Catalog Number 99-01172-00, software versions 11.1 and 11.1.1 Product Description: The Merge Hemo, Model RCSV2 device is a hemodynamics recording computer system that monitors, measures, displays, records and stores various physiologic and blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices. The system is used for documenting and monitoring patients undergoing diagnostic and interventional catheter-based procedures. The procedures can be performed in the clinical areas of invasive cardiology, interventional radiology, and cardiac electrophysiology.

    Lot or code
    Schiller Model Name: ARGUS PB-3000, Schiller Model #: 3.921002, Affected PB-3000 Serial Numbers: 7012.000035, 7012.000036, 7012.000037, 7012.000038, 7012.000041, 7012.000042, 7012.000044, 7012.000045, 7012.000046, 7012.000047, 7012.000048, 7012.000050, 7012.000051, 7012.000052, 7012.000053, 7012.000057, 7012.000058, 7012.000059, 7012.000060, 7012.000061, 7012.000062, 7012.000063, 7012.000064, 7012.000065, 7012.000066, 7012.000067, 7012.000068, 7012.000069, 7012.000071, 7012.000072, 7012.000073, 7012.000074, 7012.000075, 7012.000076, 7012.000078, 7012.000079, 7012.000080, 7012.000081, 7012.000082, 7012.000083, 7012.000085, 7012.000087, 7012.000088, 7012.000089, 7012.000090, 7012.000091, 7012.000092, 7012.000093, 7012.000094, 7012.000097, 7012.000098, 7012.000099, 7012.000100, 7012.000101, 7012.000102, 7012.000103, 7012.000104, 7012.000105, 7012.000106, 7012.000107, 7012.000108, 7012.000109, 7012.000110, 7012.000111, 7012.000112, 7012.000113, 7012.000114, 7012.000115, 7012.000116, 7012.000117, 7012.000118, 7012.000119, 7012.000120, 7012.000123, 7012.000124, 7012.000125, 7012.000126, 7012.000137, 7012.000138, 7012.000140, 7012.000141, 7012.000142, 7012.000157, 7012.000176
    FDA recall number
    Z-2518-2025
    Quantity
    84

Distribution (FDA): US Nationwide distribution including in the states of Arizona, Arkansas, California, Florida, Louisiana, Mississippi, North Carolina, North Dakota, Oklahoma, Pennsylvania, Georgia, Illinois, Kentucky, Tennessee, Texas, Washington, and Wisconsin.

Source

Issued by
U.S. Food and Drug Administration
Firm
Merge Healthcare, Inc.
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Merge Healthcare, Inc., Month: September 2025, Week 2025-W37. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Merge Healthcare has been informed by Schiller AG (Baar, Switzerland) of an issue with the Schiller ARGUS PB-3000 Patient Data Module (PDM) where certain units trigger an error message on the Merge Hemo Client (Pressure Sensor Defective) during non-invasive blood pressure (NIBP) measurements, specifically when the initial inflation pressure is set in the high-pressure range (>240 mmHg). The ARGUS PB-3000 is a vital signs monitoring instrument intended to be integrated into the Merge Hemo System. No other components of the Merge Hemo System are affected.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 97429 on September 10, 2025.

Other notices of Merge Healthcare, Inc.