Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 97416: ORTOMA AB: Ortoma Inserter Adapter OTD: Inserter Adapter OTD Straight REF 30-201 Model Number: 2001…

FDA recall 97416 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Due to a weld breakage between the pin and inserter holder of the Inserter Adapter.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

7 more codes are in the record below.

  • Ortoma Inserter Adapter OTD: Inserter Adapter OTD Straight REF 30-201 Model Number: 2001 UDI-DI code: 07350137521019 Inserter Adapter OTD Pinnacle Straight REF 30-202 Model Number: 1001 UDI-DI code: 07350137520593 Reamer Adapter OTD REF 30-157 Model Number: 1003 UDI-DI code: 07350137520081 Inserter Adapter OTD REF 30-158 Model Number: 1004 UDI-DI code: 07350137520098 The Inserter Adapters are intended to be attached to a Cup Inserter Tool. The Cup Inserter Tool (impactor) is intended to be tracked by and navigated to orient and position the hip implant cup correctly relative to preoperative planned position and orientation.

    Lot or code
    Model Number: 2001 Catalog Number: 30-201 UDI-DI code: 07350137521019 Lot Numbers: 20244803 Model Number: 1001 Catalog Number: 30-202 UDI-DI code: 07350137520593 Lot Numbers: 20245103 20245104 20245109 Model Number: 1003 Catalog Number: 30-157 UDI-DI code: 07350137520081 Lot Numbers: 20242001 20242008 20245119 Model Number: 1004 Catalog Number: 30-158 UDI-DI code: 07350137520098 Only distributed O.U.S.
    FDA recall number
    Z-2615-2025
    Quantity
    261 devices

Distribution (FDA): Worldwide - U.S. Nationwide distribution in the state of FL and the countries of Japan, Sweden.

Source

Issued by
U.S. Food and Drug Administration
Firm
ORTOMA AB
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Month: October 2025, Week 2025-W40. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Due to a weld breakage between the pin and inserter holder of the Inserter Adapter.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 97416 on October 1, 2025.