Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 97396: Boston Scientific Corporation: Boston Scientific Encore 26 Inflation Device and 3 more

FDA recall 97396 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

82 more codes are in the record below.

  • Boston Scientific Encore 26 Inflation Device, Material Numbers (UPN): 1. H74904526011; 2. H74904526052; 3. M0067101140; 4. M001151050; 5. M001151062; 6. M00566670. (Interventional Cardiology, Peripheral Interventions)

    Lot or code
    1. H74904526011, GTIN 8714729177029, Lot Numbers: 34915076, 35065978; 2. H74904526052, GTIN 8714729127062, Lot Numbers: 34915078, 34915220; 3. M0067101140, GTIN 8714729755814, Lot Numbers: 34892421; 4. M001151050, GTIN 8714729183624, Lot Numbers: 34966934; 5. M001151062, GTIN 8714729137542, Lot Numbers: 34873498; 6. M00566670, GTIN 8714729755241, Lot Numbers: 35045668.
    FDA recall number
    Z-2533-2025
    Quantity
    39739 units
  • Boston Scientific Encore Advantage Kit, Material Numbers (UPN): 1. H74904527011; 2. H74904527052. (Interventional Cardiology and Peripheral Interventions)

    Lot or code
    1. H74904527011, GTIN 08714729180005, Lot Numbers: 34783060, 34849364, 34873499, 34942522, 34976509, 35017763, 35065979, 35067000; 2. H74904527052, GTIN 08714729127048, Lot Numbers: 34873681, 34915075, 34942524, 34986473, 34986475.
    FDA recall number
    Z-2534-2025
    Quantity
    68347 units
  • Boston Scientific NephroMax Kit, Material Numbers (UPN): 1. M0062101180; 2. M0062101600. (Urology)

    Lot or code
    1. M0062101180, GTIN 08714729077589, Lot Numbers: 35275630, 35603649, 35618255, 35644223, 35768091; 2. M0062101600, GTIN 08714729834540, Lot Numbers: 35268077, 35294478, 35547681, 35576896, 35627459.
    FDA recall number
    Z-2535-2025
    Quantity
    458 units
  • Boston Scientific UroMax Ultra Kit, Material Numbers (UPN): 1. M0062251200, 2. M0062251210, 3. M0062251220, 4. M0062251230, 5. M0062251240, 6. M0062251260, 7. M0062251290, 8. M0062251300, 9. M0062251310. 10. M0062251350, 11. M0062251360, 12. M0062251370. (Urology)

    Lot or code
    1. M0062251200, GTIN 8714729341277, Lot Numbers: 35536688, 35538633, 35538634, 35543803, 35543804, 35572728, 35678100; 2. M0062251210, GTIN 8714729341284, Lot Numbers: 35274797, 35274799, 35277223, 35277224, 35294475, 35294476, 35584203, 35584215, 35656798, 35667482, 35667483; 3. M0062251220, GTIN 8714729341291, Lot Numbers: 35257438, 35535450, 35536202, 35547680, 35656799; 4. M0062251230, GTIN 8714729341307, Lot Numbers: 35266389, 35543805; 5. M0062251240, GTIN 8714729341314, Lot Numbers: 35690647; 6. M0062251260, GTIN 8714729341338, Lot Numbers: 35537838, 35537839. 7. M0062251290, GTIN 08714729341369, Lot Numbers: 35266391; 8. M0062251300, GTIN 08714729341376, Lot Numbers: 35275806, 35536992; 9. M0062251310, GTIN 08714729341383, Lot Numbers: 35536991; 10. M0062251350, GTIN 08714729341390, Lot Numbers: 35690648; 11. M0062251360, GTIN 08714729302421, Lot Numbers: 35257611, 35536990; 12. M0062251370, GTIN 08714729302438, Lot Numbers: 35294477, 35536689.
    FDA recall number
    Z-2536-2025
    Quantity
    1369 units

Distribution (FDA): Worldwide distribution.

Source

Issued by
U.S. Food and Drug Administration
Firm
Boston Scientific Corporation (other notices of this firm)
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Boston Scientific Corporation, Month: September 2025, Week 2025-W38. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 97396 on September 17, 2025.