What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
3 more codes are in the record below.
WATCHMAN TruSeal Access System SGL, OUS, Material Number (UPN) M635TS70010; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.
- Lot or code
- GTIN 08714729965732, ALL NON-EXPIRED BATCHES
- FDA recall number
- Z-2426-2025
- Quantity
- 340185 units in total
WATCHMAN TruSeal Access System DBL, OUS, Material Number (UPN) M635TS70020; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.
- Lot or code
- GTIN 08714729965749, ALL NON-EXPIRED BATCHES
- FDA recall number
- Z-2427-2025
- Quantity
- 340185 units in total
WATCHMAN TruSeal Access System ANT, OUS, Material Number (UPN) M635TS70040; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.
- Lot or code
- GTIN 08714729965756, ALL NON-EXPIRED BATCHES
- FDA recall number
- Z-2428-2025
- Quantity
- 340185 units in total
WATCHMAN FXD Curve Access Sys Sgl, OUS, Material Number (UPN) M635TS80010; intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems.
- Lot or code
- GTIN 00191506013820, ALL NON-EXPIRED BATCHES
- FDA recall number
- Z-2429-2025
- Quantity
- 340185 units in total
WATCHMAN FXD Curve Access Sys Dbl, OUS, Material Number (UPN) M635TS80020; intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems.
- Lot or code
- GTIN 00191506013837, ALL NON-EXPIRED BATCHES
- FDA recall number
- Z-2430-2025
- Quantity
- 340185 units in total
WATCHMAN TruSeal Access System SGL, US, Material Number (UPN) M635TU70010; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.
- Lot or code
- GTIN 08714729965701, ALL NON-EXPIRED BATCHES
- FDA recall number
- Z-2431-2025
- Quantity
- 340185 units in total
WATCHMAN TruSeal Access System DBL, US, Material Number (UPN) M635TU70020; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.
- Lot or code
- GTIN 08714729965718, ALL NON-EXPIRED BATCHES
- FDA recall number
- Z-2432-2025
- Quantity
- 340185 units in total
WATCHMAN TruSeal Access System ANT, US, Material Number (UPN) M635TU70040; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.
- Lot or code
- GTIN 08714729965725, ALL NON-EXPIRED BATCHES
- FDA recall number
- Z-2433-2025
- Quantity
- 340185 units in total
WATCHMAN FXD Curve Access System Sgl, US, Material Number (UPN) M635TU80010; intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems.
- Lot or code
- GTIN 00191506013806, ALL NON-EXPIRED BATCHES
- FDA recall number
- Z-2434-2025
- Quantity
- 340185 units in total
WATCHMAN FXD Curve Access System Dbl, US, Material Number (UPN) M635TU80020; intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems.
- Lot or code
- GTIN 00191506013813, ALL NON-EXPIRED BATCHES
- FDA recall number
- Z-2435-2025
- Quantity
- 340185 units in total
WATCHMAN TruSteer Access System, US, Material Number (UPN) M635TU90050;intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems.
- Lot or code
- GTIN 00191506022310, ALL NON-EXPIRED BATCHES
- FDA recall number
- Z-2436-2025
- Quantity
- 340185 units in total
Distribution (FDA): Worldwide.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Boston Scientific Corporation (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.